Pull-Out Strength of Orthodontic Miniscrews in the Temporal Bone
Authors/Creators
-
1.
Medizinische Hochschule Hannover
- 2. Otojig GmbH, Hannover, Germany
-
3.
Hearing4all
-
4.
Universitätsmedizin Göttingen
Description
Background Minimally invasive cochlear implant surgery by using a microstereotactic frame demands solid connection to the bone. We aimed to determine the stability of commercially available orthodontic miniscrews to evaluate their feasibility for frame's fixation. In addition, which substitute material most closely resembles the mechanical properties of the human temporal bone was evaluated. Methods Pull-out tests were carried out with five different types of orthodontic miniscrews in human temporal bone specimens. Furthermore, short fiber filled epoxy (SFFE), solid rigid polyurethane (SRPU50), bovine femur, and porcine iliac bone were evaluated as substitute materials. In total, 57 tests in human specimens and 180 tests in the substitute materials were performed. Results In human temporal bone, average pull-out forces ranged from 220 N to 285 N between screws. Joint stiffness in human temporal bone ranged between 14 N/mm and 358 N/mm. Statistically significant differences between the tested screws were measured in terms of stiffness and elastic energy. One screw type failed insertion due to tip breakage. No significant differences occurred between screws in maximum pull-out force. The average pull-out values of SFFE were 14.1 N higher compared to human specimen. Conclusion Orthodontic miniscrews provided rigid fixation when partially inserted in human temporal bone, as evidenced by pull-out forces and joint stiffness. Average values exceeded requirements despite variations between screws. Differences in stiffness and elastic energy indicate screw-specific interface mechanics. With proper insertion, orthodontic miniscrews appear suitable for microstereotactic frame anchoring during minimally invasive cochlear implant surgery. However, testing under more complex loading is needed to better predict clinical performance. For further pull-out tests, the most suitable substitute material is SFFE.
Additional details
Identifiers
- OpenAlex
- W4399908287
- PMID
- 38903014
- URL
- https://journals.sagepub.com/doi/10.1177/19160216241248669
Funding
-
Deutsche Forschungsgemeinschaft
- 13GW0367B
-
Deutsche Forschungsgemeinschaft
- 2177/1
-
Deutsche Forschungsgemeinschaft
- EXC2177/1
-
Deutsche Forschungsgemeinschaft
- EXC 2177: Hearing4all 390895286
-
Deutsche Forschungsgemeinschaft
- EXC2177
Dates
- Issued
-
June 21, 2024
References
- Deep, N., Dowling, E., Jethanamest, D., & Carlson, M. (2018). Cochlear Implantation: An Overview. Journal of Neurological Surgery Part B: Skull Base, 80(02), 169–177. https://doi.org/10.1055/s-0038-1669411
- https://doi.org/10.3390/jcm10194305.4305 https://doi.org/10.3390/jcm10194305.4305
- Varadarajan, V. V., Sydlowski, S. A., Li, M. M., Anne, S., & Adunka, O. F. (2020). Evolving Criteria for Adult and Pediatric Cochlear Implantation. Ear, Nose & Throat Journal, 100(1), 31–37. https://doi.org/10.1177/0145561320947258
- Sargsyan, G., Kanaan, N., Lenarz, T., & Lesinski-Schiedat, A. (2021). Comparison of speech recognition in cochlear implant patients with and without residual hearing: A review of indications. Cochlear Implants International, 22(5), 257–264. https://doi.org/10.1080/14670100.2021.1898111
- Peters, J. P. M., van Heteren, J. A. A., Wendrich, A. W., van Zanten, G. A., Grolman, W., Stokroos, R. J., Smit, A. L., & Berger, V. (2021). Short-term outcomes of cochlear implantation for single-sided deafness compared to bone conduction devices and contralateral routing of sound hearing aids—Results of a Randomised controlled trial (CINGLE-trial). PLOS ONE, 16(10), e0257447. https://doi.org/10.1371/journal.pone.0257447
- Arndt, S., Laszig, R., Aschendorff, A., Hassepass, F., Beck, R., & Wesarg, T. (2017). Cochlear implant treatment of patients with single-sided deafness or asymmetric hearing loss. HNO, 65(S2), 98–108. https://doi.org/10.1007/s00106-016-0297-5
- Carlson, M. L., Breen, J. T., Driscoll, C. L., Link, M. J., Neff, B. A., Gifford, R. H., & Beatty, C. W. (2012). Cochlear Implantation in Patients With Neurofibromatosis Type 2. Otology & Neurotology, 33(5), 853–862. https://doi.org/10.1097/mao.0b013e318254fba5
- Hansen, M. R., Gantz, B. J., & Dunn, C. (2013). Outcomes After Cochlear Implantation for Patients With Single-Sided Deafness, Including Those With Recalcitrant Ménière's Disease. Otology & Neurotology, 34(9), 1681–1687. https://doi.org/10.1097/mao.0000000000000102
- Buechner, A., Brendel, M., Lesinski-Schiedat, A., Wenzel, G., Frohne-Buechner, C., Jaeger, B., & Lenarz, T. (2010). Cochlear Implantation in Unilateral Deaf Subjects Associated With Ipsilateral Tinnitus. Otology & Neurotology, 31(9), 1381–1385. https://doi.org/10.1097/mao.0b013e3181e3d353
- Lenarz, T. (2018). Cochlear implant – state of the art. GMS Current Topics in Otorhinolaryngology - Head and Neck Surgery; 16:Doc04. https://doi.org/10.3205/cto000143
- Carlson, M. L., & Ingelfinger, J. R. (2020). Cochlear Implantation in Adults. New England Journal of Medicine, 382(16), 1531–1542. https://doi.org/10.1056/nejmra1904407
- Rau, T. S., Kreul, D., Lexow, J., Hügl, S., Zuniga, M. G., Lenarz, T., & Majdani, O. (2019). Characterizing the size of the target region for atraumatic opening of the cochlea through the facial recess. Computerized Medical Imaging and Graphics, 77, 101655. https://doi.org/10.1016/j.compmedimag.2019.101655
- Kobler, J.-P., Nuelle, K., Lexow, G. J., Rau, T. S., Majdani, O., Kahrs, L. A., Kotlarski, J., & Ortmaier, T. (2015). Configuration optimization and experimental accuracy evaluation of a bone-attached, parallel robot for skull surgery. International Journal of Computer Assisted Radiology and Surgery, 11(3), 421–436. https://doi.org/10.1007/s11548-015-1300-4
- Caversaccio, M., Wimmer, W., Anso, J., Mantokoudis, G., Gerber, N., Rathgeb, C., Schneider, D., Hermann, J., Wagner, F., Scheidegger, O., Huth, M., Anschuetz, L., Kompis, M., Williamson, T., Bell, B., Gavaghan, K., Weber, S., & Pérez-Fernández, N. (2019). Robotic middle ear access for cochlear implantation: First in man. PLOS ONE, 14(8), e0220543. https://doi.org/10.1371/journal.pone.0220543
- Rau, T. S., Witte, S., Uhlenbusch, L., Kahrs, L. A., Lenarz, T., & Majdani, O. (2021). Concept description and accuracy evaluation of a moldable surgical targeting system. Journal of Medical Imaging, 8(01). https://doi.org/10.1117/1.jmi.8.1.015003
- Labadie, R. F., Balachandran, R., Noble, J. H., Blachon, G. S., Mitchell, J. E., Reda, F. A., Dawant, B. M., & Fitzpatrick, J. M. (2014). Minimally invasive image‐guided cochlear implantation surgery: First report of clinical implementation. The Laryngoscope, 124(8), 1915–1922. https://doi.org/10.1002/lary.24520
- Salcher, R., John, S., Stieghorst, J., Kluge, M., Repp, F., Fröhlich, M., & Lenarz, T. (2022). Minimally Invasive Cochlear Implantation: First-in-Man of Patient-Specific Positioning Jigs. Frontiers in Neurology, 13. https://doi.org/10.3389/fneur.2022.829478
- Kobler, J.-P., Kotlarski, J., Lexow, G. J., Majdani, O., & Ortmaier, T. (2014). Design optimization of a bone-attached, redundant and reconfigurable parallel kinematic device for skull surgery. 2014 IEEE International Conference on Robotics and Automation (ICRA), 2364–2371. https://doi.org/10.1109/icra.2014.6907187
- Rau, T. S., John, S., Kluge, M., Repp, F., Zuniga, M. G., Stieghorst, J., Timm, M. E., Frohlich, M., Majdani, O., & Lenarz, T. (2023). Ex Vivo Evaluation of a Minimally Invasive Approach for Cochlear Implant Surgery. IEEE Transactions on Biomedical Engineering, 70(1), 390–398. https://doi.org/10.1109/tbme.2022.3192144
- Labadie, R. F., Balachandran, R., Mitchell, J. E., Noble, J. H., Majdani, O., Haynes, D. S., Bennett, M. L., Dawant, B. M., & Fitzpatrick, J. M. (2010). Clinical Validation Study of Percutaneous Cochlear Access Using Patient-Customized Microstereotactic Frames. Otology & Neurotology, 31(1), 94–99. https://doi.org/10.1097/mao.0b013e3181c2f81a
- Müller, S., Kahrs, L. A., Gaa, J., Tauscher, S., Kluge, M., John, S., Rau, T. S., Lenarz, T., Ortmaier, T., & Majdani, O. (2019). Workflow assessment as a preclinical development tool. International Journal of Computer Assisted Radiology and Surgery, 14(8), 1389–1401. https://doi.org/10.1007/s11548-019-02002-3
- Labadie, R. F., Mitchell, J., Balachandran, R., & Fitzpatrick, J. M. (2009). Customized, rapid-production microstereotactic table for surgical targeting: description of concept and in vitro validation. International Journal of Computer Assisted Radiology and Surgery, 4(3), 273–280. https://doi.org/10.1007/s11548-009-0292-3
- Kobler, J.-P., Prielozny, L., Lexow, G. J., Rau, T. S., Majdani, O., & Ortmaier, T. (2015). Mechanical characterization of bone anchors used with a bone-attached, parallel robot for skull surgery. Medical Engineering & Physics, 37(5), 460–468. https://doi.org/10.1016/j.medengphy.2015.02.012
- Teekavanich, C., Uezono, M., Takakuda, K., Ogasawara, T., Techalertpaisarn, P., & Moriyama, K. (2021). Evaluation of Cortical Bone Microdamage and Primary Stability of Orthodontic Miniscrew Using a Human Bone Analogue. Materials, 14(8), 1825. https://doi.org/10.3390/ma14081825
- Nenen, F., Garnica, N., Rojas, V., & Oyonarte, R. (2020). Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles:. The Angle Orthodontist, 91(3), 336–342. https://doi.org/10.2319/050120-375.1
- Pithon, M. M., Nojima, M. G., & Nojima, L. I. (2012). Primary stability of orthodontic mini-implants inserted into maxilla and mandible of swine. Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 113(6), 748–754. https://doi.org/10.1016/j.tripleo.2011.06.021
- Florvaag, B., Kneuertz, P., Lazar, F., Koebke, J., Zöller, J. E., Braumann, B., & Mischkowski, R. A. (2010). Biomechanical Properties of Orthodontic Miniscrews. An In-vitro Study. Journal of Orofacial Orthopedics / Fortschritte Der Kieferorthopädie, 71(1), 53–67. https://doi.org/10.1007/s00056-010-9933-y
- Gracco, A., Giagnorio, C., Incerti Parenti, S., Alessandri Bonetti, G., & Siciliani, G. (2012). Effects of thread shape on the pullout strength of miniscrews. American Journal of Orthodontics and Dentofacial Orthopedics, 142(2), 186–190. https://doi.org/10.1016/j.ajodo.2012.03.023
- Yashwant, A. V. (2017). Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro Study. JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH. https://doi.org/10.7860/jcdr/2017/25774.9808
- Schwartz‐Dabney, C. L., & Dechow, P. C. (2003). Variations in cortical material properties throughout the human dentate mandible. American Journal of Physical Anthropology, 120(3), 252–277. https://doi.org/10.1002/ajpa.10121
- Delille, R., Lesueur, D., Potier, P., Drazetic, P., & Markiewicz, E. (2007). Experimental study of the bone behaviour of the human skull bone for the development of a physical head model. International Journal of Crashworthiness, 12(2), 101–108. https://doi.org/10.1080/13588260701433081
- Motherway, J. A., Verschueren, P., Van der Perre, G., Vander Sloten, J., & Gilchrist, M. D. (2009). The mechanical properties of cranial bone: The effect of loading rate and cranial sampling position. Journal of Biomechanics, 42(13), 2129–2135. https://doi.org/10.1016/j.jbiomech.2009.05.030
- Ayers, M. P., Clift, S. E., & Gheduzzi, S. (2013). Morsellised sawbones is an acceptable experimental substitute for the in vitro elastic and viscoelastic mechanical characterisation of morsellised cancellous bone undergoing impaction grafting. Medical Engineering & Physics, 36(1), 26–31. https://doi.org/10.1016/j.medengphy.2013.08.005
- Reilly, D. T., & Burstein, A. H. (1975). The elastic and ultimate properties of compact bone tissue. Journal of Biomechanics, 8(6), 393–405. https://doi.org/10.1016/0021-9290(75)90075-5
- Willis-Owen, C. A., Hearn, T. C., Keene, G. C., & Costi, J. J. (2010). Biomechanical testing of implant free wedge shaped bone block fixation for bone patellar tendon bone anterior cruciate ligament reconstruction in a bovine model. Journal of Orthopaedic Surgery and Research, 5(1), 66. https://doi.org/10.1186/1749-799x-5-66
- Shaw, J. M., Hunter, S. A., Gayton, C. J., Boivin, G. P., & Prayson, M. J. (2012). Repeated Freeze-thaw Cycles Do Not Alter the Biomechanical Properties of Fibular Allograft Bone. Clinical Orthopaedics & Related Research, 470(3), 937–943. https://doi.org/10.1007/s11999-011-2033-5
- Sawbone Europe. Biomechanical test materials properties. Accessed August 15, 2021. https://www.sawbone.com
- Peterson, J., & Dechow, P. C. (2003). Material properties of the human cranial vault and zygoma. The Anatomical Record Part A: Discoveries in Molecular, Cellular, and Evolutionary Biology, 274A(1), 785–797. https://doi.org/10.1002/ar.a.10096
- Brosh, T., Rozitsky, D., Geron, S., Pilo, R., & Glogauer, M. (2014). Tensile Mechanical Properties of Swine Cortical Mandibular Bone. PLoS ONE, 9(12), e113229. https://doi.org/10.1371/journal.pone.0113229
- Migliorati, M., Drago, S., Dalessandri, D., Lagazzo, A., Gallo, F., Capurro, M., & Silvestrini-Biavati, A. (2018). On the stability efficiency of anchorage self-tapping screws: Ex vivo experiments on miniscrew implants used in orthodontics. Journal of the Mechanical Behavior of Biomedical Materials, 81, 46–51. https://doi.org/10.1016/j.jmbbm.2018.02.019
- Kobler, J.-P., Wall, S., Lexow, G. J., Lang, C. P., Majdani, O., Kahrs, L. A., & Ortmaier, T. (2015). An experimental evaluation of loads occurring during guided drilling for cochlear implantation. International Journal of Computer Assisted Radiology and Surgery, 10(10), 1625–1637. https://doi.org/10.1007/s11548-015-1153-x
- Kratchman, L. B., Blachon, G. S., Withrow, T. J., Balachandran, R., Labadie, R. F., & Webster, R. J. (2011). Design of a Bone-Attached Parallel Robot for Percutaneous Cochlear Implantation. IEEE Transactions on Biomedical Engineering, 58(10), 2904–2910. https://doi.org/10.1109/tbme.2011.2162512
- Rau, T. S., Zuniga, M. G., Salcher, R., & Lenarz, T. (2020). A simple tool to automate the insertion process in cochlear implant surgery. International Journal of Computer Assisted Radiology and Surgery, 15(11), 1931–1939. https://doi.org/10.1007/s11548-020-02243-7
- Talon, E., Visini, M., Wagner, F., Caversaccio, M., & Wimmer, W. (2021). Quantitative Analysis of Temporal Bone Density and Thickness for Robotic Ear Surgery. Frontiers in Surgery, 8. https://doi.org/10.3389/fsurg.2021.740008