Introduction Impedance telemetry measurements in cochlear implant (CI) recipients are commonly used to assess CI electrode functionality and provide valuable insights into inner ear conditions. However, these measurements usually take place only during surgery and at clinical follow-up appointments, offering limited temporal resolution of impedance changes. This study aimed to address this gap by implementing daily impedance monitoring using a smartphone app. Methods A prospective study evaluated the usability of a research app for remote impedance measurements over 4 months following standard CI surgery with a MED-EL FLEX28 electrode. Impedance was recorded twice daily (morning and evening). The mean impedance across all electrode channels was analyzed for four postoperative time periods: early postoperative (up to day 10 postoperatively), late postoperative (from day 11 to ~4 weeks), intensive fitting (from ~4 weeks postoperatively to ~7 weeks) and regular hearing phase (from ~7 weeks to 4 months). Two CI fitting approaches were compared: activation during the early postoperative phase (early activation, EA) and activation during the intensive fitting phase (conventional activation, CA). Morning-to-evening differences in impedance (MED) were also examined. Results The app demonstrated an overall usage rate of 66% ( n = 28), indicating moderate-to-high adherence. Except for higher evening impedance values with CA in the late postoperative phase, no significant differences in mean impedance between the fitting approaches were observed (EA: 6.46 kΩ, n = 11; CA: 7.82 kΩ, n = 11; p = 0.04). Significant differences in MED were found during the early postoperative phase (EA: 0.06 kΩ, n = 8; CA: −0.18 kΩ, n = 10; p = 0.04) and the late postoperative phases (EA: 0.85 kΩ, n = 11; CA: 0.03 kΩ, n = 11; p < 0.001). Conclusion Remote impedance measurements via the app can be made over an extended postoperative period. The increased measurement frequency allowed for detailed characterization of impedance dynamics, particularly around the onset of electrical stimulation. No clinically relevant difference in mean impedance was found between EA and CA groups. Daily fluctuations showed consistently lower evening values after stimulation onset. These findings highlight the potential value of this approach for enhancing postoperative CI management.

Additional details

Dates

Issued
July 4, 2025

References

  1. Gstoettner, W., Kiefer, J., Baumgartner, W.-D., Pok, S., Peters, S., & Adunka, O. (2004). Hearing preservation in cochlear implantation for electric acoustic stimulation. Acta Oto-Laryngologica, 124(4), 348–352. https://doi.org/10.1080/00016480410016432
  2. Helbig, S., Adel, Y., Rader, T., Stöver, T., & Baumann, U. (2016). Long-term Hearing Preservation Outcomes After Cochlear Implantation for Electric-Acoustic Stimulation. Otology & Neurotology, 37(9), e353–e359. https://doi.org/10.1097/mao.0000000000001066
  3. James, C., Albegger, K., Battmer, R., Burdo, S., Deggouj, N., Deguine, O., Dillier, N., Gersdorff, M., Laszig, R., Lenarz, T., Rodriguez, M. M., Mondain, M., Offeciers, E., Macías, Á. R., Ramsden, R., Sterkers, O., Von Wallenberg, E., Weber, B., & Fraysse, B. (2005). Preservation of residual hearing with cochlear implantation: How and why. Acta Oto-Laryngologica, 125(5), 481–491. https://doi.org/10.1080/00016480510026197
  4. Thompson, N. J., Dillon, M. T., Buss, E., Park, L. R., Pillsbury, H. C., O'Connell, B. P., & Brown, K. D. (2020). Electrode Array Type and Its Impact on Impedance Fluctuations and Loss of Residual Hearing in Cochlear Implantation. Otology & Neurotology, 41(2), 186–191. https://doi.org/10.1097/mao.0000000000002457
  5. Choi, J., Payne, M. R., Campbell, L. J., Bester, C. W., Newbold, C., Eastwood, H., & O'Leary, S. J. (2017). Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation. Otology & Neurotology, 38(10), 1433–1439. https://doi.org/10.1097/mao.0000000000001589
  6. Wimmer, W., Sclabas, L., Caversaccio, M., & Weder, S. (2022). Cochlear Implant Electrode Impedance as Potential Biomarker for Residual Hearing. Frontiers in Neurology, 13. https://doi.org/10.3389/fneur.2022.886171
  7. Shaul, C., Bester, C. W., Weder, S., Choi, J., Eastwood, H., Padmavathi, K. V., Collins, A., & O'Leary, S. J. (2019). Electrical Impedance as a Biomarker for Inner Ear Pathology Following Lateral Wall and Peri-modiolar Cochlear Implantation. Otology & Neurotology, 40(5), e518–e526. https://doi.org/10.1097/mao.0000000000002227
  8. Foggia, M. J., Quevedo, R. V., & Hansen, M. R. (2019). Intracochlear fibrosis and the foreign body response to cochlear implant biomaterials. Laryngoscope Investigative Otolaryngology, 4(6), 678–683. https://doi.org/10.1002/lio2.329
  9. Tykocinski, M., Cohen, L. T., & Cowan, R. S. (2005). Measurement and Analysis of Access Resistance and Polarization Impedance in Cochlear Implant Recipients. Otology & Neurotology, 26(5), 948–956. https://doi.org/10.1097/01.mao.0000185056.99888.f3
  10. Newbold, C., Richardson, R., Millard, R., Huang, C., Milojevic, D., Shepherd, R., & Cowan, R. (2010). Changes in biphasic electrode impedance with protein adsorption and cell growth. Journal of Neural Engineering, 7(5), 056011. https://doi.org/10.1088/1741-2560/7/5/056011
  11. Newbold, C., Richardson, R., Millard, R., Seligman, P., Cowan, R., & Shepherd, R. (2011). Electrical stimulation causes rapid changes in electrode impedance of cell-covered electrodes. Journal of Neural Engineering, 8(3), 036029. https://doi.org/10.1088/1741-2560/8/3/036029
  12. Newbold, C., Richardson, R., Huang, C. Q., Milojevic, D., Cowan, R., & Shepherd, R. (2004). Anin vitromodel for investigating impedance changes with cell growth and electrical stimulation: implications for cochlear implants. Journal of Neural Engineering, 1(4), 218–227. https://doi.org/10.1088/1741-2560/1/4/005
  13. Newbold, C., Mergen, S., Richardson, R., Seligman, P., Millard, R., Cowan, R., & Shepherd, R. (2013). Impedance changes in chronically implanted and stimulated cochlear implant electrodes. Cochlear Implants International, 15(4), 191–199. https://doi.org/10.1179/1754762813y.0000000050
  14. Wei, J. J.-F., Tung, T.-H., Li, L. P.-H., & Buechner, A. (2021). Evolution of impedance values in cochlear implant patients after early switch-on. PLOS ONE, 16(2), e0246545. https://doi.org/10.1371/journal.pone.0246545
  15. Hu, H.-C., Chen, J. K.-C., Tsai, C.-M., Chen, H.-Y., Tung, T.-H., Li, L. P.-H., & Zeng, F.-G. (2017). Evolution of impedance field telemetry after one day of activation in cochlear implant recipients. PLOS ONE, 12(3), e0173367. https://doi.org/10.1371/journal.pone.0173367
  16. Di Lella, F. A., De Marco, D., Fernández, F., Parreño, M., & Boccio, C. M. (2019). In Vivo Real-time Remote Cochlear Implant Capacitive Impedance Measurements: A Glimpse Into the Implanted Inner Ear. Otology & Neurotology, 40(5S), S18–S22. https://doi.org/10.1097/mao.0000000000002214
  17. Nadol, J. B., Burgess, B. J., Gantz, B. J., Coker, N. J., Ketten, D. R., Kos, I., Roland, J. T., Shiao, J. Y., Eddington, D. K., Montandon, P., & Shallop, J. K. (2001). Histopathology of Cochlear Implants in Humans. Annals of Otology, Rhinology & Laryngology, 110(9), 883–891. https://doi.org/10.1177/000348940111000914
  18. Benatti, A., Castiglione, A., Trevisi, P., Bovo, R., Rosignoli, M., Manara, R., & Martini, A. (2013). Endocochlear inflammation in cochlear implant users: Case report and literature review. International Journal of Pediatric Otorhinolaryngology, 77(6), 885–893. https://doi.org/10.1016/j.ijporl.2013.03.016
  19. Araki, S., Kawano, A., Lee Seldon, H., Shepherd, R. K., Funasaka, S., Clark, G. M., & Myers, E. N. (2000). Effects of intracochlear factors on spiral ganglion cells and auditory brain stem response after long‐term electrical stimulation in deafened kittens. Otolaryngology–Head and Neck Surgery, 122(3), 425–433. https://doi.org/10.1067/mhn.2000.93863
  20. Wilk, M., Hessler, R., Mugridge, K., Jolly, C., Fehr, M., Lenarz, T., Scheper, V., & Yamamoto, M. (2016). Impedance Changes and Fibrous Tissue Growth after Cochlear Implantation Are Correlated and Can Be Reduced Using a Dexamethasone Eluting Electrode. PLOS ONE, 11(2), e0147552. https://doi.org/10.1371/journal.pone.0147552
  21. Prenzler, N. K., Salcher, R., Lenarz, T., Gaertner, L., & Warnecke, A. (2020). Dose-Dependent Transient Decrease of Impedances by Deep Intracochlear Injection of Triamcinolone With a Cochlear Catheter Prior to Cochlear Implantation–1 Year Data. Frontiers in Neurology, 11. https://doi.org/10.3389/fneur.2020.00258
  22. Paasche, G., Tasche, C., Stöver, T., Lesinski-Schiedat, A., & Lenarz, T. (2009). The Long-Term Effects of Modified Electrode Surfaces and Intracochlear Corticosteroids on Postoperative Impedances in Cochlear Implant Patients. Otology & Neurotology, 30(5), 592–598. https://doi.org/10.1097/mao.0b013e3181ab8fba
  23. Briggs, R., O ’Leary, S., Birman, C., Plant, K., English, R., Dawson, P., Risi, F., Gavrilis, J., Needham, K., & Cowan, R. (2020). Comparison of electrode impedance measures between a dexamethasone-eluting and standard Cochlear™ Contour Advance® electrode in adult cochlear implant recipients. Hearing Research, 390, Article 107924. https://doi.org/10.1016/j.heares.2020.107924
  24. Prenzler, N., Salcher, R., Büchner, A., Warnecke, A., Kley, D., Batsoulis, C., Vormelcher, S., Mitterberger-Vogt, M., Morettini, S., Schilp, S., Hochmair, I., & Lenarz, T. (2025). Cochlear implantation with a dexamethasone-eluting electrode array: First-in-human safety and performance results. Hearing Research, 461, Article 109255. https://doi.org/10.1016/j.heares.2025.109255
  25. Hughes, M. L., Vander Werff, K. R., Brown, C. J., Abbas, P. J., Kelsay, D. M. R., Teagle, H. F. B., & Lowder, M. W. (2001). A Longitudinal Study of Electrode Impedance, the Electrically Evoked Compound Action Potential, and Behavioral Measures in Nucleus 24 Cochlear Implant Users. Ear and Hearing, 22(6), 471–486. https://doi.org/10.1097/00003446-200112000-00004

1–25 of 47