Oversensing in a Cardioverter Defibrillator System Caused by Interaction Between Two Endocardial Defibrillation Leads in the Right Ventricle

Description

A 53‐year‐old male patient underwent implantable Cardioverter defibrillator (ICD) implantation with a single lead (Endotak ®) transvenous system due to recurrent episodes of drug refractory ventricular tachycardia. After pulse generator replacement, inappropriate ICD shocks were observed due to muscle potential sensing. Intraoperatively, the old Endotak ® lead could not be extracted; therefore, it was transsected and capped. A new lead was inserted and tested without any problems. At the predischarge test, VF was induced and was followed by ICD shocks during sinus rhythm. In another surgical procedure, the old Endotak ® lead was explanted using a special instrument. The present report demonstrates that two endocardial Endotak ® leads should be avoided, because the leads may disturb each other and be followed by inappropriate ICD discharges.

Additional details

Dates

Issued
April 1998

References

  1. Mirowski, M., Reid, P. R., Mower, M. M., Watkins, L., Gott, V. L., Schauble, J. F., Langer, A., Heilman, M. S., Kolenik, S. A., Fischell, R. E., & Weisfeldt, M. L. (1980). Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human Beings. New England Journal of Medicine, 303(6), 322–324. https://doi.org/10.1056/nejm198008073030607
  2. EDEL, T. B., MALONEY, J. D., MOORE, S., MCGUINN, W. P., GOHN, D., WILLIAMS, D., WILKOFF, B. L., MCCARTHY, P., MORANT, V. A., PEARCE, G., SIMMONS, T. W., TROHMAN, R., FIRSTENBERG, M., & CASTLE, L. W. (1991). Six‐Year Clinical Experience with the Automatic Implantable Cardioverter Defibrillator. Pacing and Clinical Electrophysiology, 14(11), 1850–1854. https://doi.org/10.1111/j.1540-8159.1991.tb02778.x
  3. TRAPPE, H., KLEIN, H., FIEGUTH, H., KIELBLOCK, B., WENZLAFF, P., & LICHTLEN, P. R. (1993). Initial Experience with a New Transvenous Defibrillation System. Pacing and Clinical Electrophysiology, 16(1), 134–140. https://doi.org/10.1111/j.1540-8159.1993.tb01550.x
  4. BLOCK, M., HAMMEL, D., ISBRUCH, F., BORGGREFE, M., WIETHOLT, D., HACHENBERG, T., SCHELD, H. H., & BREITHARDT, G. (1992). Results and Realistic Expectations with Transvenous Lead Systems. Pacing and Clinical Electrophysiology, 15(4), 665–670. https://doi.org/10.1111/j.1540-8159.1992.tb05160.x
  5. GRIMM, W., FLORES, B. F., & MARCHLINSKI, F. E. (1993). Complications of Implantable Cardioverter Defibrillator Therapy: Follow‐Up of 241 Patients. Pacing and Clinical Electrophysiology, 16(1), 218–222. https://doi.org/10.1111/j.1540-8159.1993.tb01565.x
  6. Almassi, G. H., Olinger, G. N., Wetherbee, J. N., & Fehl, G. (1993). Long-term complications of implantable cardioverter defibrillator lead systems. The Annals of Thoracic Surgery, 55(4), 888–892. https://doi.org/10.1016/0003-4975(93)90112-u
  7. PETERS, R. W., FOSTER, A. H., SHOROFSKY, S. R., NOLAN, D. A., & GOLD, M. R. (1995). Spurious Discharges Due to Late Insulation Break in Endocardial Sensing Leads for Cardioverter Defibrillators. Pacing and Clinical Electrophysiology, 18(3), 478–481. https://doi.org/10.1111/j.1540-8159.1995.tb02549.x
  8. Stambler, B. S., Wood, M. A., Damiano, R. J., Greenway, P. S., Smutka, M. L., & Ellenbogen, K. A. (1994). Sensing/pacing lead complications with a newer generation implantable cardioverter-defibrillator: Worldwide experience from the guardian ATP 4210 clinical trial. Journal of the American College of Cardiology, 23(1), 123–132. https://doi.org/10.1016/0735-1097(94)90510-x
  9. HORTON, R. P., CANBY, R. C., ROMÁN, C. A., JESSEN, M. E., SCHULTE, T. J., & PAGE, R. L. (1995). Diagnosis of ICD Lead Failure Using Continuous Event Marker Recording. Pacing and Clinical Electrophysiology, 18(6), 1331–1334. https://doi.org/10.1111/j.1540-8159.1995.tb06976.x
  10. BALLAS, S. L., RASHIDI, R., MCALISTER, H., CORBELLI, R., MCCOWAN, R., WILKOFF, B. L., CASTLE, L. W., MORANT, V. A., SIMMONS, T. W., & MALONEY, J. D. (1989). The Use of Beep‐O‐Grams in the Assessment of Automatic Implantahle Cardioverter Defibrillator Sensing Function. Pacing and Clinical Electrophysiology, 12(11), 1737–1745. https://doi.org/10.1111/j.1540-8159.1989.tb01859.x
  11. A new instrument for the removal of infected cardiac pacemaker electrodes

Citations

  1. Cha, Y.-M., El-Chami, M. F., Liu, C. F., Andreychuk, L. J., Beaver, T. M., Bergen, K. M., Berul, C. I., Birgersdotter-Green, U. M., Breitenstein, A., Epstein, L. M., Gross, J. N., Jackson, L. R., Karim, S., Krahn, A. D., Kusumoto, F., Lever, N., Linton-Frazier, L. N., Love, C. J., Mah, D. Y., … Zeitler, E. (2026). 2026 HRS/AHA/APHRS/EHRA/IDSA/LAHRS/PACES/STS expert consensus statement update on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm, 23(8), e1716–e1792. https://doi.org/10.1016/j.hrthm.2026.04.015
  2. Gładysz-Wańha, S., Joniec, M., Wańha, W., Piłat, E., Drzewiecka, A., Gardas, R., Biernat, J., Węglarzy, A., & Gołba, K. S. (2024). Transvenous lead extraction safety and efficacy in infected and noninfected patients using mechanical-only tools: Prospective registry from a high-volume center. Heart Rhythm, 21(4), 427–435. https://doi.org/10.1016/j.hrthm.2023.12.015
  3. Miguelena Hycka, J., López Menéndez, J., Martín García, M., Muñoz Pérez, R., Castro Pinto, M., Torres Terreros, C. B., García Chumbiray, P. F., & Rodriguez-Roda, J. (2023). Electrodos no funcionantes ¿Extracción o abandono?. Cirugía Cardiovascular, 30(4), 251–258. https://doi.org/10.1016/j.circv.2022.11.005
  4. Teixeira, R. A., Fagundes, A. A., Baggio, J. M., Oliveira, J. C. de ., Medeiros, P. de T. J., Valdigem, B. P., Teno, L. A. C., Silva, R. T., Melo, C. S. de ., Elias, J., Moraes, A. V., Pedrosa, A. A. A., Porto, F. M., Brito, H. L. de ., Souza, T. G. S. e ., Mateos, J. C. P., Moraes, L. G. B. de ., Forno, A. R. J. D., D'Avila, A. L. B., … Leal, J. C. F. (2023). Diretriz Brasileira de Dispositivos Cardíacos Eletrônicos Implantáveis – 2023. Arquivos Brasileiros De Cardiologia, 120(1). https://doi.org/10.36660/abc.20220892
  5. https://doi.org/10.3389/fphys.2022.993604
  6. Swerdlow, C. D., Friedman, P. A., Asirvatham, S. J., & Hayes, D. L. (2021, April 19). Troubleshooting. In Cardiac Pacing, Defibrillation and Resynchronization (1st ed., pp. 399–513). Wiley. https://doi.org/10.1002/9781119264002.ch10
  7. Sidhu, B. S., Gould, J., Sieniewicz, B., Porter, B., & Rinaldi, C. A. (2018). The role of transvenous lead extraction in the management of redundant or malfunctioning pacemaker and defibrillator leads post ELECTRa. EP Europace. https://doi.org/10.1093/europace/euy018
  8. Swerdlow, C. D., Brown, M. L., & Bordachar, P. (2017). Sensing and Detection With Cardiac Implantable Electronic Devices. In Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy (pp. 114–167). Elsevier. https://doi.org/10.1016/b978-0-323-37804-8.00004-3
  9. Kusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R., Cha, Y.-M., Clancy, J., Deharo, J.-C., Ellenbogen, K. A., Exner, D., Hussein, A. A., Kennergren, C., Krahn, A., Lee, R., Love, C. J., Madden, R. A., Mazzetti, H. A., Moore, J. C., … Wazni, O. (2017). 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm, 14(12), e503–e551. https://doi.org/10.1016/j.hrthm.2017.09.001
  10. Swerdlow, C. D., Asirvatham, S. J., Ellenbogen, K. A., & Friedman, P. A. (2014). Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I. Circulation: Arrhythmia and Electrophysiology, 7(6), 1237–1261. https://doi.org/10.1161/circep.114.002344
  11. Verlato, R., Baccillieri, M. S., & Turrini, P. (2011). From Guidelines: Definitions, Indications, Facilities, and Outcomes of Transvenous Lead Extraction. In Transvenous Lead Extraction (pp. 33–45). Springer Milan. https://doi.org/10.1007/978-88-470-1466-4_2
  12. Amelot, M., Foucault, A., Scanu, P., Gomes, S., Champ-Rigot, L., Pellissier, A., & Milliez, P. (2011). Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter defibrillator leads. Archives of Cardiovascular Diseases, 104(11), 572–577. https://doi.org/10.1016/j.acvd.2011.08.004
  13. Glikson, M., Suleiman, M., Luria, D. M., Martin, M. L., Hodge, D. O., Shen, W.-K., Bradley, D. J., Munger, T. M., Rea, R. F., Hayes, D. L., Hammill, S. C., & Friedman, P. A. (2009). Do abandoned leads pose risk to implantable cardioverter-defibrillator patients?. Heart Rhythm, 6(1), 65–68. https://doi.org/10.1016/j.hrthm.2008.10.012
  14. Wilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., Epstein, L. M., Friedman, R. A., Kennergren, C. E. H., Mitkowski, P., Schaerf, R. H. M., & Wazni, O. M. (2009). Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient Management. Heart Rhythm, 6(7), 1085–1104. https://doi.org/10.1016/j.hrthm.2009.05.020
  15. Glikson, M., Luria, D., Gurevitz, O., & Friedman, P. A. (n.d.). Beyond Sudden Death Prevention: Minimizing Implantable Cardioverter Defibrillator Shocks and Morbidity and Optimizing Efficacy. In Electrical Diseases of the Heart (pp. 781–800). Springer London. https://doi.org/10.1007/978-1-84628-854-8_56
  16. Cooper, J. M., Sauer, W. H., Garcia, F. C., Krautkramer, M. J., & Verdino, R. J. (2007). Covering sleeves can shield the high-voltage coils from lead chatter in an integrated bipolar ICD lead. EP Europace, 9(2), 137–142. https://doi.org/10.1093/europace/eul180
  17. https://doi.org/10.1111/j.1540-8167.2007.00843.x
  18. https://doi.org/10.1111/j.1540-8159.2006.00510.x
  19. Erlinger, P. J., & Bardy, G. H. (2004, February 17). Interpretation of Device Stored Rhythms and Electrocardiograms. In Dynamic Electrocardiography (1st ed., pp. 608–625). Wiley. https://doi.org/10.1002/9780470987483.ch64
  20. MARRAZZO, N., UGOLOTTI, U., VILLANI, G. Q., GROPPI, F., CASTELLINI, P., & CAPUCCI, A. (2003). Endocardial Defibrillation Lead Extraction:. Pacing and Clinical Electrophysiology, 26(10), 2036–2038. https://doi.org/10.1046/j.1460-9592.2003.00314.x
  21. Russo, A. M., & Marchlinski, F. E. (2002). Long-Term Follow-Up. In Developments in Cardiovascular Medicine (pp. 161–243). Springer US. https://doi.org/10.1007/978-1-4615-1055-0_9
  22. Meine, M., Pfitzner, P., Voigt, B., & Trappe, H.-J. (2001). Akute Herzrhythmusstörungen bei Schrittmacher- und Defibrillatorpatienten. In Die Notfalltherapie bei akuten Herzrhythmusstörungen (pp. 51–68). Steinkopff. https://doi.org/10.1007/978-3-642-57631-7_6