Evaluating different methods of MR-based motion correction in simultaneous PET/MR using a head phantom moved by a robotic system

Description

Abstract Background Due to comparatively long measurement times in simultaneous positron emission tomography and magnetic resonance (PET/MR) imaging, patient movement during the measurement can be challenging. This leads to artifacts which have a negative impact on the visual assessment and quantitative validity of the image data and, in the worst case, can lead to misinterpretations. Simultaneous PET/MR systems allow the MR-based registration of movements and enable correction of the PET data. To assess the effectiveness of motion correction methods, it is necessary to carry out measurements on phantoms that are moved in a reproducible way. This study explores the possibility of using such a phantom-based setup to evaluate motion correction strategies in PET/MR of the human head. Method An MR-compatible robotic system was used to generate rigid movements of a head-like phantom. Different tools, either from the manufacturer or open-source software, were used to estimate and correct for motion based on the PET data itself (SIRF with SPM and NiftyReg) and MR data acquired simultaneously (e.g. MCLFIRT, BrainCompass). Different motion estimates were compared using data acquired during robot-induced motion. The effectiveness of motion correction of PET data was evaluated by determining the segmented volume of an activity-filled flask inside the phantom. In addition, the segmented volume was used to determine the centre-of-mass and the change in maximum activity concentration. Results The results showed a volume increase between 2.7 and 36.3% could be induced by the experimental setup depending on the motion pattern. Both, BrainCompass and MCFLIRT, produced corrected PET images, by reducing the volume increase to 0.7–4.7% (BrainCompass) and to -2.8–0.4% (MCFLIRT). The same was observed for example for the centre-of-mass, where the results show that MCFLIRT (0.2–0.6 mm after motion correction) had a smaller deviation from the reference position than BrainCompass (0.5–1.8 mm) for all displacements. Conclusions The experimental setup is suitable for the reproducible generation of movement patterns. Using open-source software for motion correction is a viable alternative to the vendor-provided motion-correction software.

Additional details

Funding

Strahlenschutzseminar in Thüringen e.V.
ROR icon Siemens Healthineers (Germany)
F2015-04 or C00229543
UK EPSRC Grant
EP/M022587/1
EPSRC Grant
EP/P022200/1
BMBF Grant
FKZ: 03Z1L511
Universitätsklinikum Magdeburg
ROR icon Bundesministerium für Forschung, Technologie und Raumfahrt
Zentrum für Innovationskompetenz - Verbundprojekt: SONO-RAY - Kombinationstherapie für Tumoren mit MR-geführtem fokussiertem Ultraschall und Strahlentherapie, TP: In-Silico-Modelle der FUS-Wirkung, Validierung in digitalen Patientenmodellen und Integration in den klinischen Workflow 03Z1L511
ROR icon Engineering and Physical Sciences Research Council
CCP in Synergistic Reconstruction for Biomedical Imaging EP/T026693/1

Dates

Issued
March 3, 2022

References

  1. Cherry, S. R., Louie, A. Y., & Jacobs, R. E. (2008). The Integration of Positron Emission Tomography With Magnetic Resonance Imaging. Proceedings of the IEEE, 96(3), 416–438. https://doi.org/10.1109/jproc.2007.913502
  2. Catana, C. (2017). Principles of Simultaneous PET/MR Imaging. Magnetic Resonance Imaging Clinics of North America, 25(2), 231–243. https://doi.org/10.1016/j.mric.2017.01.002
  3. Gillman, A., Smith, J., Thomas, P., Rose, S., & Dowson, N. (2017). PET motion correction in context of integrated PET/MR: Current techniques, limitations, and future projections. Medical Physics, 44(12). https://doi.org/10.1002/mp.12577
  4. Fürst, S., Grimm, R., Hong, I., Souvatzoglou, M., Casey, M. E., Schwaiger, M., Nekolla, S. G., & Ziegler, S. I. (2015). Motion Correction Strategies for Integrated PET/MR. Journal of Nuclear Medicine, 56(2), 261–269. https://doi.org/10.2967/jnumed.114.146787
  5. Catana, C., Benner, T., van der Kouwe, A., Byars, L., Hamm, M., Chonde, D. B., Michel, C. J., El Fakhri, G., Schmand, M., & Sorensen, A. G. (2010). MRI-Assisted PET Motion Correction for Neurologic Studies in an Integrated MR-PET Scanner. Journal of Nuclear Medicine, 52(1), 154–161. https://doi.org/10.2967/jnumed.110.079343
  6. Chen, K. T., Izquierdo-Garcia, D., Poynton, C. B., Chonde, D. B., & Catana, C. (2016). On the accuracy and reproducibility of a novel probabilistic atlas-based generation for calculation of head attenuation maps on integrated PET/MR scanners. European Journal of Nuclear Medicine and Molecular Imaging, 44(3), 398–407. https://doi.org/10.1007/s00259-016-3489-z
  7. Rahmim, A., Rousset, O., & Zaidi, H. (2007). Strategies for Motion Tracking and Correction in PET. PET Clinics, 2(2), 251–266. https://doi.org/10.1016/j.cpet.2007.08.002
  8. Bloomfield, P. M., Spinks, T. J., Reed, J., Schnorr, L., Westrip, A. M., Livieratos, L., Fulton, R., & Jones, T. (2003). The design and implementation of a motion correction scheme for neurological PET. Physics in Medicine and Biology, 48(8), 959–978. https://doi.org/10.1088/0031-9155/48/8/301
  9. Mukherjee, J. M., Lindsay, C., Mukherjee, A., Olivier, P., Shao, L., King, M. A., & Licho, R. (2016). Improved frame-based estimation of head motion in PET brain imaging. Medical Physics, 43(5), 2443–2454. https://doi.org/10.1118/1.4946814
  10. Noonan, P. J., Howard, J., Hallett, W. A., & Gunn, R. N. (2015). Repurposing the Microsoft Kinect for Windows v2 for external head motion tracking for brain PET. Physics in Medicine and Biology, 60(22), 8753–8766. https://doi.org/10.1088/0031-9155/60/22/8753
  11. Jiao, J., Searle, G. E., Schnabel, J. A., & Gunn, R. N. (2015). Impact of image-based motion correction on dopamine D3/D2 receptor occupancy—comparison of groupwise and frame-by-frame registration approaches. EJNMMI Physics, 2(1). https://doi.org/10.1186/s40658-015-0117-0
  12. Olesen, O. V., Sullivan, J. M., Mulnix, T., Paulsen, R. R., Hojgaard, L., Roed, B., Carson, R. E., Morris, E. D., & Larsen, R. (2013). List-Mode PET Motion Correction Using Markerless Head Tracking: Proof-of-Concept With Scans of Human Subject. IEEE Transactions on Medical Imaging, 32(2), 200–209. https://doi.org/10.1109/tmi.2012.2219693
  13. Olesen, O. V., Paulsen, R. R., Hojgaard, L., Roed, B., & Larsen, R. (2012). Motion Tracking for Medical Imaging: A Nonvisible Structured Light Tracking Approach. IEEE Transactions on Medical Imaging, 31(1), 79–87. https://doi.org/10.1109/tmi.2011.2165157
  14. Matsubara, K., Ibaraki, M., Nakamura, K., Yamaguchi, H., Umetsu, A., Kinoshita, F., & Kinoshita, T. (2013). Impact of subject head motion on quantitative brain 15O PET and its correction by image-based registration algorithm. Annals of Nuclear Medicine, 27(4), 335–345. https://doi.org/10.1007/s12149-013-0690-z
  15. Inomata, T., Watanuki, S., Odagiri, H., Nambu, T., Karakatsanis, N. A., Ito, H., Watabe, H., Tashiro, M., & Shidahara, M. (2019). A systematic performance evaluation of head motion correction techniques for 3 commercial PET scanners using a reproducible experimental acquisition protocol. Annals of Nuclear Medicine, 33(7), 459–470. https://doi.org/10.1007/s12149-019-01353-w
  16. Kyme, A. Z., & Fulton, R. R. (2021). Motion estimation and correction in SPECT, PET and CT. Physics in Medicine & Biology, 66(18), 18TR02. https://doi.org/10.1088/1361-6560/ac093b
  17. Spangler‐Bickell, M. G., Hurley, S. A., Deller, T. W., Jansen, F., Bettinardi, V., Carlson, M., Zeineh, M., Zaharchuk, G., & McMillan, A. B. (2021). Optimizing the frame duration for data‐driven rigid motion estimation in brain PET imaging. Medical Physics, 48(6), 3031–3041. https://doi.org/10.1002/mp.14889
  18. Polycarpou, I., Soultanidis, G., & Tsoumpas, C. (2021). Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging. Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, 379(2204). https://doi.org/10.1098/rsta.2020.0207
  19. Herzog H, Tellmann L, Fulton R, Stangier I, Kops ER, Bente K, Boy C, Hurlemann R, Pietrzyk U. Motion artifact reduction on parametric pet images of neuroreceptor binding. J Nucl Med. 2005;46(6):1059–65.
  20. Ullisch, M. G., Scheins, J. J., Weirich, C., Rota Kops, E., Celik, A., Tellmann, L., Stöcker, T., Herzog, H., Shah, N. J., & Chen, C.-T. (2012). MR-Based PET Motion Correction Procedure for Simultaneous MR-PET Neuroimaging of Human Brain. PLoS ONE, 7(11), e48149. https://doi.org/10.1371/journal.pone.0048149
  21. Catana, C., Drzezga, A., Heiss, W.-D., & Rosen, B. R. (2012). PET/MRI for Neurologic Applications. Journal of Nuclear Medicine, 53(12), 1916–1925. https://doi.org/10.2967/jnumed.112.105346
  22. Grimm, R., Fürst, S., Souvatzoglou, M., Forman, C., Hutter, J., Dregely, I., Ziegler, S. I., Kiefer, B., Hornegger, J., Block, K. T., & Nekolla, S. G. (2015). Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI. Medical Image Analysis, 19(1), 110–120. https://doi.org/10.1016/j.media.2014.08.003
  23. Ikari, Y., Nishio, T., Makishi, Y., Miya, Y., Ito, K., Koeppe, R. A., & Senda, M. (2012). Head motion evaluation and correction for PET scans with 18F-FDG in the Japanese Alzheimer's disease neuroimaging initiative (J-ADNI) multi-center study. Annals of Nuclear Medicine, 26(7), 535–544. https://doi.org/10.1007/s12149-012-0605-4
  24. Ashburner, J., & Friston, K. (2007). Rigid Body Registration. In Statistical Parametric Mapping (pp. 49–62). Elsevier. https://doi.org/10.1016/b978-012372560-8/50004-8
  25. Mourik, J. E. M., Lubberink, M., van Velden, F. H. P., Lammertsma, A. A., & Boellaard, R. (2009). Off-line motion correction methods for multi-frame PET data. European Journal of Nuclear Medicine and Molecular Imaging, 36(12), 2002–2013. https://doi.org/10.1007/s00259-009-1193-y

1–25 of 41