Published December 29, 2025 | Journal article | Version v1 | https://doi.org/10.1007/s10544-025-00766-x

Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load

Description

Abstract Roughly 5% of the population is affected by hearing loss. In case of a functional middle ear impairment with decreased transmission, different types of active middle ear implants may restore hearing. To drive the round window membrane (RWM) for example, bone anchored actuators or floating mass transducer (FMT) have been used in clinics. Bone anchored devices are usually too big for the round window niche and FMTs display low displacement and force at low frequencies. This paper focuses on the stimulation of the RWM by a stator-based actuator that combines small size and bone anchoring. The design features a rotational-symmetric balanced armature core with flexible membrane parts manufactured micro technologically. Membranes and assembled prototypes were evaluated on the bench in the range of 0.1 kHz – 10 kHz and prototypes tested in human temporal bones. We demonstrate the feasibility of a balanced armature actuator of appropriate size for RW placement. With 1.4 mm outer diameter and 4.9 mm length, the actuator fits the anatomical constraints. Acoustic stimulation of the RWM resulted in a maximum output of approximately 80 eq. dB SPL at 300 Hz increasing to > 100 eq. dB SPL above 5 kHz. The measured output is comparable to clinically used devices. The design lacked sufficient robustness leading to a low yield in manufacturing and failures in experiments, although the output level was insensitive to static force loading up to 30 mN. Future developments require an increased maximum output level and improved robustness. Graphical abstract

Files

s10544-025-00766-x.pdf

Files (1.9 MB)

Name Size Download all
md5:2b22ac89657f8e987484e889a94a9bf2
1.9 MB Preview Download

Additional details

Dates

Issued
December 29, 2025

References

  1. ANSI. American national standard s Institute Methods for Calculation of the Speech Intelligibility Index . (ANSI S3.5–1997) (1997)
  2. ASTM International, Standard practice for describing system output of implantable middle ear hearing devices. ASTM F2504–05, 1–6 (2005)
  3. Beltrame, A. M., Martini, A., Prosser, S., Giarbini, N., & Streitberger, C. (2009). Coupling the Vibrant Soundbridge to Cochlea Round Window: Auditory Results in Patients With Mixed Hearing Loss. Otology & Neurotology, 30(2), 194–201. https://doi.org/10.1097/mao.0b013e318180a495
  4. A. Glukhovskoy, E. Mueller, M.S. Prediger, H. Maier, M.C. Wurz, Development of a polyimid membrane for electromagnetic middle- ear actuators. MikroSystemTechnik Congress 2021, Proceedings. VDE Verlag GmbH, S. 186-189 (2021)
  5. Heckeler, C., & Eiber, A. (2017). Mechanical Aspects of the Round Window Stimulation. Procedia IUTAM, 24, 15–29. https://doi.org/10.1016/j.piutam.2017.08.039
  6. Maier, H., Salcher, R., Schwab, B., & Lenarz, T. (2013). The effect of static force on round window stimulation with the direct acoustic cochlea stimulator. Hearing Research, 301, 115–124. https://doi.org/10.1016/j.heares.2012.12.010
  7. Maier, H., Lenarz, T., & Busch, S. (2024). An easy method to determine crucial AMEI performance parameters from clinical routine data in individuals - Part 1: maximum output. International Journal of Audiology, 64(3), 290–297. https://doi.org/10.1080/14992027.2024.2341100
  8. Müller, M., Salcher, R., Lenarz, T., & Maier, H. (2017). The Hannover Coupler: Controlled Static Prestress in Round Window Stimulation With the Floating Mass Transducer. Otology & Neurotology, 38(8), 1186–1192. https://doi.org/10.1097/mao.0000000000001484
  9. Rosowski, J. J., Chien, W., Ravicz, M. E., & Merchant, S. N. (2007). Testing a Method for Quantifying the Output of Implantable Middle Ear Hearing Devices. Audiology and Neurotology, 12(4), 265–276. https://doi.org/10.1159/000101474
  10. Salcher, R., Schwab, B., Lenarz, T., & Maier, H. (2014). Round window stimulation with the floating mass transducer at constant pretension. Hearing Research, 314, 1–9. https://doi.org/10.1016/j.heares.2014.04.001
  11. Schraven, S. P., Hirt, B., Goll, E., Heyd, A., Gummer, A. W., Zenner, H.-P., & Dalhoff, E. (2011). Conditions for Highly Efficient and Reproducible Round-Window Stimulation in Humans. Audiology and Neurotology, 17(2), 133–138. https://doi.org/10.1159/000333807
  12. P. Schreier, Simulation based design of new implantable hearing device. COMSOL News. Cochlear Technology Centre Belgium, Mechelen, Belgium (2012)
  13. Shin, D. H., Lee, K.-Y., & Seong, K. W. (2024). An acoustic bellows-type round window transducer for middle-ear implants. Clinical Biomechanics, 120, 106349. https://doi.org/10.1016/j.clinbiomech.2024.106349
  14. V. Vacuumschmelze, VACUUMSCHMELZE GMBH & CO. KG: Weichmagnetische Werkstoffe und Halbzeuge.PERMENORM H2/V5. (2024)
  15. [if gte mso 9]>[endif][if gte mso 9] Normal 0 21 false false false DE X-NONE X-NONE [endif][if gte mso 9][endif][if gte mso 10][endif] https://www.who.int/news-room/fact-sheets/detail/deafness-and-hearing-loss
  16. V. Colletti, M. Carner, L. Colletti, L. Sacchetto and N. Giarbini (2006). "Round window mechanical stimulation for severe mixed hearing loss." Association for Research in Otolaryngology Abstracts 29; 987.
  17. Knölke, N., Murawski, D., Wardenga, N., Busch, S., Maier, H., & Lenarz, T. (2022). The Hannover Coupler V2: Audiological outcomes of a round window coupler for the Vibrant Soundbridge. Laryngoscope Investigative Otolaryngology, 7(3), 840–846. https://doi.org/10.1002/lio2.812
  18. Häusler, R., Stieger, C., Bernhard, H., & Kompis, M. (2008). A Novel Implantable Hearing System with Direct Acoustic Cochlear Stimulation. Audiology and Neurotology, 13(4), 247–256. https://doi.org/10.1159/000115434
  19. Iwasaki, S., Suzuki, H., Moteki, H., Miyagawa, M., Takumi, Y., & Usami, S.-. ichi . (2012). Experience with the Vibrant Soundbridge RW-Coupler for round window Vibroplasty with tympanosclerosis. Acta Oto-Laryngologica, 132(6), 676–682. https://doi.org/10.3109/00016489.2011.649492
  20. Kiefer, J., Arnold, W., & Staudenmaier, R. (2006). Round Window Stimulation with an Implantable Hearing Aid (Soundbridge®) Combined with Autogenous Reconstruction of the Auricle – A New Approach. ORL, 68(6), 378–385. https://doi.org/10.1159/000095282
  21. Wollenberg, B., Beltrame, M., Schönweiler, R., Gehrking, E., Nitsch, S., Steffen, A., & Frenzel, H. (2007). Integration des aktiven Mittelohrimplantates in die plastische Ohrmuschelrekonstruktion. HNO, 55(5), 349–356. https://doi.org/10.1007/s00106-007-1540-x
  22. Lefebvre, P. P., Martin, C., Dubreuil, C., Decat, M., Yazbeck, A., Kasic, J., & Tringali, S. (2008). A Pilot Study of the Safety and Performance of the Otologics Fully Implantable Hearing Device: Transducing Sounds via the Round Window Membrane to the Inner Ear. Audiology and Neurotology, 14(3), 172–180. https://doi.org/10.1159/000171479