Published August 27, 2025 | Journal article | Version v1 | https://doi.org/10.1177/19714009251373070

Safety and efficacy of the pipeline vantage flow diverter for the treatment of saccular aneurysms in the posterior cerebral circulation

Description

Introduction Posterior circulation aneurysms are particularly challenging to treat due to their anatomical complexity and the high perforator density within this region. The pipeline vantage flow diverter (PVFD) has shown promising results in treating anterior circulation aneurysms. However, its efficacy and safety in treating posterior circulation saccular aneurysms are not well investigated. Methods Single-center study reviewed patients with posterior circulation aneurysms treated with the PVFD between September 2021 and March 2024. Patients and aneurysm characteristics, clinical results, and radiological results were documented. Results 22 patients harboring 24 aneurysms were identified. All aneurysms had a saccular morphology. Complications included ischemic events in two patients (8.3%), one leading to mRS deterioration from 0 to 3. One cerebral hemorrhage leading to mRS shift from 0 to 1 was also documented. At the latest imaging follow-up available (median 14.3 months), complete aneurysm occlusion (Class I, Raymond-Roy occlusion classification (RROC) was achieved in 50% of the cases, residual neck (Class II) in 41.7%, and residual aneurysm (Class III) in 8.3%. Basilar bifurcation aneurysms had lower complete occlusion rates (RROC I: 33.3%) and higher complication rates (16.7% with mRS shift) compared to other locations. In-stent stenosis was rare (4.5%). A limitation of the study is the retrospective, single-center study design. Conclusion The PVFD demonstrates high occlusion rates and a favorable safety profile in the treatment of saccular aneurysms in the posterior circulation. However, treatment of basilar bifurcation aneurysms with the PVFD remains challenging due to the complex anatomy and high-flow dynamics in this location.

Additional details

Dates

Issued
August 27, 2025

References

  1. Peluso, J. P., van Rooij, W. J., Sluzewski, M., Beute, G. N., & Majoie, C. B. (2007). Posterior Inferior Cerebellar Artery Aneurysms: Incidence, Clinical Presentation, and Outcome of Endovascular Treatment. American Journal of Neuroradiology, 29(1), 86–90. https://doi.org/10.3174/ajnr.a0758
  2. Williamson, R. W., Wilson, D. A., Abla, A. A., McDougall, C. G., Nakaji, P., Albuquerque, F. C., & Spetzler, R. F. (2015). Clinical characteristics and long-term outcomes in patients with ruptured posterior inferior cerebellar artery aneurysms: a comparative analysis. Journal of Neurosurgery, 123(2), 441–445. https://doi.org/10.3171/2014.10.jns141079
  3. Unruptured Intracranial Aneurysms — Risk of Rupture and Risks of Surgical Intervention. (1998). Unruptured Intracranial Aneurysms — Risk of Rupture and Risks of Surgical Intervention. New England Journal of Medicine, 339(24), 1725–1733. https://doi.org/10.1056/nejm199812103392401
  4. Tjahjadi, M., Serrone, J., & Hernesniemi, J. (2018). Should we still consider clips for basilar apex aneurysms? A critical appraisal of the literature. Surgical Neurology International, 9(1), 44. https://doi.org/10.4103/sni.sni_311_17
  5. Wiebers, D. O. (2003). Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. The Lancet, 362(9378), 103–110. https://doi.org/10.1016/s0140-6736(03)13860-3
  6. Al-Afif, S., Lang, J. M., Abdulbaki, A., Palmaers, T., Scheinichen, D., Abu-Fares, O., Hermann, E. J., & Krauss, J. K. (2024). The safety and utility of the semi-sitting position for clipping of posterior circulation aneurysms. Acta Neurochirurgica, 166(1), Article 341. https://doi.org/10.1007/s00701-024-06229-1
  7. Bohnstedt, B. N., Ziemba-Davis, M., Sethia, R., Payner, T. D., DeNardo, A., Scott, J., & Cohen-Gadol, A. A. (2017). Comparison of endovascular and microsurgical management of 208 basilar apex aneurysms. Journal of Neurosurgery, 127(6), 1342–1352. https://doi.org/10.3171/2016.8.jns16703
  8. Adeeb, N., Ogilvy, C. S., Griessenauer, C. J., & Thomas, A. J. (2019). Expanding the Indications for Flow Diversion: Treatment of Posterior Circulation Aneurysms. Neurosurgery, 86(Supplement_1), S76–S84. https://doi.org/10.1093/neuros/nyz344
  9. Griessenauer, C. J., Ogilvy, C. S., Adeeb, N., Dmytriw, A. A., Foreman, P. M., Shallwani, H., Limbucci, N., Mangiafico, S., Kumar, A., Michelozzi, C., Krings, T., Pereira, V. M., Matouk, C. C., Harrigan, M. R., Shakir, H. J., Siddiqui, A. H., Levy, E. I., Renieri, L., Marotta, T. R., … Thomas, A. J. (2019). Pipeline embolization of posterior circulation aneurysms: a multicenter study of 131 aneurysms. Journal of Neurosurgery, 130(3), 923–935. https://doi.org/10.3171/2017.9.jns171376
  10. Hellstern, V., Aguilar-Pérez, M., Henkes, E., Serna-Candel, C., Wendl, C., Bäzner, H., Ganslandt, O., & Henkes, H. (2021). Endovascular Treatment of Posterior Circulation Saccular Aneurysms With the p64 Flow Modulation Device: Mid-and Long-Term Results in 54 Aneurysms From a Single Center. Frontiers in Neurology, 12. https://doi.org/10.3389/fneur.2021.711863
  11. Becske, T., Potts, M. B., Shapiro, M., Kallmes, D. F., Brinjikji, W., Saatci, I., McDougall, C. G., Szikora, I., Lanzino, G., Moran, C. J., Woo, H. H., Lopes, D. K., Berez, A. L., Cher, D. J., Siddiqui, A. H., Levy, E. I., Albuquerque, F. C., Fiorella, D. J., Berentei, Z., … Nelson, P. K. (2017). Pipeline for uncoilable or failed aneurysms: 3-year follow-up results. Journal of Neurosurgery, 127(1), 81–88. https://doi.org/10.3171/2015.6.jns15311
  12. Döring, K., Aburub, A., Krauss, J. K., Lang, J. M., Al-Afif, S., Polemikos, M., Weissenborn, K., Grosse, G., Grieb, D., Lanfermann, H., Götz, F., & Abu-Fares, O. (2023). Early clinical experience with the new generation Pipeline Vantage flow diverter in the treatment of unruptured saccular aneurysms using short-term dual antiplatelet therapy. Interventional Neuroradiology, 32(3), 359–367. https://doi.org/10.1177/15910199231205047
  13. King, R. M., Brooks, O. W., Langan, E. T., Caroff, J., Clarençon, F., Tamura, T., Wainwright, J. M., Gounis, M. J., Marosfői, M., & Puri, A. S. (2017). Communicating malapposition of flow diverters assessed with optical coherence tomography correlates with delayed aneurysm occlusion. Journal of NeuroInterventional Surgery, 10(7), 697–702. https://doi.org/10.1136/neurintsurg-2017-013502
  14. Martínez-Galdámez, M., Lamin, S. M., Lagios, K. G., Liebig, T., Ciceri, E. F., Chapot, R., Stockx, L., Chavda, S., Kabbasch, C., Faragò, G., Nordmeyer, H., Boulanger, T., Piano, M., & Boccardi, E. P. (2018). Treatment of intracranial aneurysms using the pipeline flex embolization device with shield technology: angiographic and safety outcomes at 1-year follow-up. Journal of NeuroInterventional Surgery, 11(4), 396–399. https://doi.org/10.1136/neurintsurg-2018-014204
  15. Trivelato, F. P., Wajnberg, E., Rezende, M. T. S., Ulhôa, A. C., Piske, R. L., Abud, T. G., de Castro-Afonso, L. H., Abath, C. G. C., Nakiri, G. S., Araújo, J. F. S., Silva, J. L. J., Tosello, R. T., Vanzin, J. R., Manzato, L. B., Baccin, C. E., da Mota, B. A. A., & Abud, D. G. (2019). Safety and Effectiveness of the Pipeline Flex Embolization Device With Shield Technology for the Treatment of Intracranial Aneurysms: Midterm Results From a Multicenter Study. Neurosurgery, 87(1), 104–111. https://doi.org/10.1093/neuros/nyz356
  16. Griessenauer, C. J., Goren, O., Dalal, S. S., & Schirmer, C. M. (2018). Pipeline Embolization Device with Shield Technology for Intracranial Aneurysms: An Initial U.S. Experience. World Neurosurgery, 119, 10–14. https://doi.org/10.1016/j.wneu.2018.07.203
  17. de Villiers, L., Carraro do Nascimento, V., Domitrovic, L., Dhillon, P. S., & Rice, H. (2024). Vanguard Study: Initial experience with the new fourth generation Pipeline Vantage Flow Diverter (PVFD): 6-month results, technical and clinical considerations. Journal of NeuroInterventional Surgery, 17(e1), e166–e171. https://doi.org/10.1136/jnis-2023-021182
  18. Booth, T. C., Bassiouny, A., Lynch, J., Sonwalkar, H., Bleakley, A., Iqbal, A., Minett, T., Buwanabala, J., Narata, A. P., Patankar, T., Islim, F. I., Kandasamy, N., Balasundaram, P., Sciacca, S., Siddiqui, J., Walsh, D., Tolias, C., Kailaya-Vasan, A., Sultan, A. A., … Sastry, A. (2023). Outcome study of the Pipeline Vantage Embolization Device (second version) in unruptured (and ruptured) aneurysms (PEDVU(R) study). Journal of NeuroInterventional Surgery, 16(11), 1136–1144. https://doi.org/10.1136/jnis-2023-020754
  19. Cortese, J., Ghozy, S., Valizadeh, P., Hasanzadeh, A., Amoukhteh, M., Jannatdoust, P., Hassankhani, A., Ahmadzade, A., Popica, D. A., Kallmes, D. F., & Kadirvel, R. (2025). Initial Experience with the Pipeline Vantage Flow Diverter for Intracranial Aneurysms: A Systematic Review and Meta-Analysis. American Journal of Neuroradiology, 46(3), 510–516. https://doi.org/10.3174/ajnr.a8555
  20. Goertz, L., Hohenstatt, S., Zopfs, D., Kottlors, J., Pennig, L., Schlamann, M., Michael, A. E., Liebig, T., A. Möhlenbruch, M., & Kabbasch, C. (2024). Pipeline Vantage Embolization Device for the treatment of intracranial aneurysms: A systematic review and meta-analysis. Interventional Neuroradiology, 32(4), 1680–1690. https://doi.org/10.1177/15910199241264340
  21. Vollherbst, D. F., Cekirge, H. S., Saatci, I., Baltacioglu, F., Onal, B., Koc, O., Rautio, R., Sinisalo, M., Tomasello, A., Vega, P., Martínez-Galdámez, M., Lynch, J., Mendes Pereira, V., Bendszus, M., & Möhlenbruch, M. A. (2022). First clinical multicenter experience with the new Pipeline Vantage flow diverter. Journal of NeuroInterventional Surgery, 15(1), 63–69. https://doi.org/10.1136/neurintsurg-2021-018480
  22. Goertz, L., Zopfs, D., Kottlors, J., Pennig, L., Schob, S., Schlamann, M., & Kabbasch, C. (2024). Treatment of intracranial aneurysms with large-diameter (≥5.5 mm) Derivo Embolization Devices, with particular focus on 7 and 8 mm diameter devices. Interventional Neuroradiology. https://doi.org/10.1177/15910199241248479
  23. Kashkoush, A., El-Abtah, M. E., Petitt, J. C., Glauser, G., Winkelman, R., Achey, R. L., Davison, M., Abdulrazzak, M. A., Hussain, S. M., Toth, G., Bain, M., & Moore, N. (2023). Flow diversion for the treatment of intracranial bifurcation aneurysms: a systematic review and meta-analysis. Journal of NeuroInterventional Surgery, 16(9), 921–927. https://doi.org/10.1136/jnis-2023-020582
  24. Sciacca, S., Bassiouny, A., Mansoor, N., Minett, T., Balasundaram, P., Siddiqui, J., Joshi, Y., Derakhshani, S., Kandasamy, N., Booth, T. C., & Lynch, J. (2023). Early Outcomes of the Pipeline Vantage Flow Diverter: A Multicentre Study. Clinical Neuroradiology, 33(4), 887–896. https://doi.org/10.1007/s00062-023-01314-x