Published October 13, 1998 | Journal article | Version v1 | https://doi.org/10.1515/cclm.1998.134

Determination of pH in Human Erythrocytes. Sources of Systematic Error

Description

Abstract The results of a simultaneous determination of the pH value in erythrocytes with the potentiometric measuring method and the 14C-labelled 5,5-dimethyl-2,4-oxazolidinedione (DMO) method showed a mean method difference of 0.026 pH units. The cause of this discrepancy was assumed to be matrix-inherent liquid-junction potentials in the potentiometric measurement. Taking these into account in the calculation leads to consistent values for the methods investigated. The DMO method proved to be free of systematic errors. Another indication of this is that its mean ratio of extracellular to intracellular H+ ion concentration (He +/Hi +) substantially agreed with the distribution ratios of other freely diffusible ions and their pH dependence.

Additional details

Dates

Issued
October 13, 1998

References

  1. Resnick, L. M., Gupta, R. K., Sosa, R. E., Corbett, M. L., & Laragh, J. H. (1987). Intracellular pH in human and experimental hypertension. Proceedings of the National Academy of Sciences, 84(21), 7663–7667. https://doi.org/10.1073/pnas.84.21.7663
  2. Fitzsimons, E. J., & Sendroy, J. (1961). Distribution of Electrolytes in Human Blood. Journal of Biological Chemistry, 236(5), 1595–1601. https://doi.org/10.1016/s0021-9258(18)64218-8
  3. Funder, J., & Wieth, J. O. (1966). Chloride and Hydrogen Ion Distribution between Human Red Cells and Plasma. Acta Physiologica Scandinavica, 68(2), 234–245. https://doi.org/10.1111/j.1748-1716.1966.tb03423.x
  4. Davis, F. E., Kenyon, K., & Kirk, J. (1953). A Rapid Titrimetric Method for Determining the Water Content of Human Blood. Science, 118(3062), 276–277. https://doi.org/10.1126/science.118.3062.276
  5. Hutchinson, E. (1952). The behavior of human erythrocytes in aqueous alcohol solutions. Archives of Biochemistry and Biophysics, 38(1), 35–41. https://doi.org/10.1016/0003-9861(52)90006-4
  6. Gary-Bobo, C. M., & Solomon, A. K. (1968). Properties of Hemoglobin Solutions in Red Cells. The Journal of General Physiology, 52(5), 825–853. https://doi.org/10.1085/jgp.52.5.825
  7. Maas, A. H. J. (1970). pH Determination of body fluids with a micro glass electrode and a saturated KCl bridge in the cell. Clinica Chimica Acta, 28(3), 373–390. https://doi.org/10.1016/0009-8981(70)90062-8
  8. Henderson, P. (1907). Zur Thermodynamik der Flüssigkeitsketten. Zeitschrift Für Physikalische Chemie, 59U(1), 118–127. https://doi.org/10.1515/zpch-1907-5907
  9. Henderson, F. (1908). Zur Thermodynamik der Flüssigkeitsketten. Zeitschrift Für Physikalische Chemie, 63U(1), 325–345. https://doi.org/10.1515/zpch-1908-6320
  10. Salling, N., & Siggaard-Andersen, O. (1971). Liquid-Junction Potentials between Plasma or Erythrolysate and KCl Solutions. Scandinavian Journal of Clinical and Laboratory Investigation, 28(1), 33–40. https://doi.org/10.3109/00365517109090660
  11. Thomason, R. (1963). The use of 5,5-dimethyl-2,4-oxazolidinedione for determination of intracellular pH. Scandinavian Journal of Clinical and Laboratory Investigation, 15(1), 45–51. https://doi.org/10.3109/00365516309051308
  12. Huot, S. J., & Aronson, P. S. (1991). Na+-H+ Exchanger and Its Role in Essential Hypertension and Diabetes Mellitus. Diabetes Care, 14(6), 521–535. https://doi.org/10.2337/diacare.14.6.521
  13. Andersen, O. S. (1961). Factors Affecting the Liouid-Junotion Potential in Electrometric Blood pH Measurement. Scandinavian Journal of Clinical and Laboratory Investigation, 13(2), 205–211. https://doi.org/10.3109/00365516109137270