Published February 6, 2026 | Journal article | Version v1 | https://doi.org/10.1177/11297298251415110

Unexpected scarring following use of capillary self-blood collection device: A case report and discussion of device design

  • 1. Department of General Practice, University Medical Center Göttingen, Göttingen, Germany
  • 2. ROR icon Michigan State University
  • 3. ROR icon Universitätsmedizin Göttingen
  • 4. Department of Clinical Chemistry, University Medical Center Göttingen, Göttingen, Germany
  • 5. Interdisciplinary UMG-Laboratories and Department of Hematology and Medical Oncology, University Medical Center Göttingen, Göttingen, Germany
  • 6. Department of Oral and Maxillofacial Surgery, University Medical Center Göttingen, Göttingen, Germany
  • 7. ROR icon Medizinische Hochschule Hannover

Description

Self-administered capillary blood collection devices are increasingly used in both clinical research and patient care. The Tasso+ device, a CE-certified system designed to collect 500 µL of whole blood, is generally regarded as safe and user-friendly. However, complications related to its use remain insufficiently documented. We describe the case of a female study participant who developed an unexpected wound complication after using the self-blood collection device. Although the blood collection itself was successful, the participant later required surgical treatment for a gaping lesion at the application site, which ultimately left a small permanent scar. To our knowledge, this represents a previously unreported adverse effect of the device. The case raises concerns about the interaction between device application and skin tension lines and underscores the need for further evaluation of safety aspects in self-administered blood collection systems.

Additional details

Funding

Europäischer Sozialfonds
ZAM 5-87006761

Dates

Issued
February 6, 2026

References

  1. Thangavelu, M. U., Wouters, B., Kindt, A., Reiss, I. K. M., & Hankemeier, T. (2023). Blood microsampling technologies: Innovations and applications in 2022. Analytical Science Advances, 4(5-6), 154–180. https://doi.org/10.1002/ansa.202300011
  2. Schröder, D., Müller, F., Heesen, G., Hummers, E., Dopfer-Jablonka, A., Vahldiek, K., Klawonn, F., Steffens, S., Mikuteit, M., Niewolik, J., Heinemann, S., & Arteaga-Livias, K. (2023). Feasibility of self-organized blood sample collection in adults for study purposes in a primary care setting. PLOS ONE, 18(5), e0286014. https://doi.org/10.1371/journal.pone.0286014
  3. Schröder, D., Hafke, A., Hummers, E., Schuchardt, C., Müller, C. A., Hoffmeister, L., Happle, C., Dopfer-Jablonka, A., Behrens, G. M. N., Schanz, J., & Müller, F. (2025). Comparison of laboratory results and pain perception in self-sampled capillary blood versus venous blood sampling: a systematic review and meta-analysis. Clinical Biochemistry, 138, Article 110965. https://doi.org/10.1016/j.clinbiochem.2025.110965
  4. Schuchardt, C., Müller, F., Hafke, A., Hummers, E., Schanz, J., Dopfer-Jablonka, A., Behrens, G. M. N., & Schröder, D. (2025). Pain and feasibility of capillary self-blood collection in general practice: A cross-sectional investigative study. European Journal of General Practice, 31(1). https://doi.org/10.1080/13814788.2025.2501309
  5. Zarbl, J., Eimer, E., Gigg, C., Bendzuck, G., Korinth, M., Elling-Audersch, C., Kleyer, A., Simon, D., Boeltz, S., Krusche, M., Mucke, J., Muehlensiepen, F., Vuillerme, N., Krönke, G., Schett, G., & Knitza, J. (2022). Remote self-collection of capillary blood using upper arm devices for autoantibody analysis in patients with immune-mediated inflammatory rheumatic diseases. RMD Open, 8(2), e002641. https://doi.org/10.1136/rmdopen-2022-002641
  6. Poland, D. C. W., & Cobbaert, C. M. (2024). Blood self-sampling devices: innovation, interpretation and implementation in total lab automation. Clinical Chemistry and Laboratory Medicine (CCLM), 63(1), 3–13. https://doi.org/10.1515/cclm-2024-0508
  7. Schmetzer, C., Vogt, E., Stellar, L., Godonou, E.-T., Liphardt, A.-M., Muehlensiepen, F., Vuillerme, N., Hueber, A. J., Kleyer, A., Krönke, G., Schett, G., Simon, D., & Knitza, J. (2022). Self-collection of capillary blood and saliva to determine COVID-19 vaccine immunogenicity in patients with immune-mediated inflammatory diseases and health professionals. Frontiers in Public Health, 10. https://doi.org/10.3389/fpubh.2022.994770
  8. Paul, S. (2018). Biodynamic excisional skin tension lines for surgical excisions: untangling the science. The Annals of The Royal College of Surgeons of England, 100(4), 330–337. https://doi.org/10.1308/rcsann.2018.0038
  9. Sommerlad, B. C., & Creasey, J. M. (1978). The stretched scar: A clinical and histological study. British Journal of Plastic Surgery, 31(1), 34–45. https://doi.org/10.1016/0007-1226(78)90012-7
  10. Blicharz, T. M., Gong, P., Bunner, B. M., Chu, L. L., Leonard, K. M., Wakefield, J. A., Williams, R. E., Dadgar, M., Tagliabue, C. A., El Khaja, R., Marlin, S. L., Haghgooie, R., Davis, S. P., Chickering, D. E., & Bernstein, H. (2018). Microneedle-based device for the one-step painless collection of capillary blood samples. Nature Biomedical Engineering, 2(3), 151–157. https://doi.org/10.1038/s41551-018-0194-1