Clinical Efficacy and Safety of the New Cardioverter Defibrillator Systems
Description
Clinical efficacy and safety of two new third‐generation implantable cardioverter defibrillators (ICD) were studied in 38 patients with ventricular tachycardia (VT) or fibrillation (VF). There were 31 patients with coronary disease, three patients with right ventricular dysplasia, one patient with dilated cardiomyopathy, and three patients with valvular disease. Twenty‐four patients (group I) received an ICD with monophasic (Ventak PRx 1700, CPI) and 14 patients (group II) with biphasic shocks (Cadence V 100, Ventritex). Intraoperatively, the mean defibrillation threshold was significantly lower in group II than in group I, both in patients with induced VT (group I 11.0 ± 6.3 joules: group N 5.8 ± 1.3 joules) (P < O.01) and induced VF (group I 17.5 ± 4.6 joules; group II 9.6 ± 5.2 joules) (P < O.O1). During the mean follow‐up of 12 ± 7 months four patients (11%) died. 865 arrhythmia events (AE) occurred and were terminated by ATP (671 VTs, 78%). Acceleration of VTs was observed in 28 AE (3%) and ATP was unable to interrupt 58 AE (7%). ICD shocks were delivered as a first therapy in 108 AE (13%).
Additional details
Identifiers
Dates
- Issued
-
January 1993
References
- Mirowski, M., Reid, P. R., Mower, M. M., Watkins, L., Gott, V. L., Schauble, J. F., Langer, A., Heilman, M. S., Kolenik, S. A., Fischell, R. E., & Weisfeldt, M. L. (1980). Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human Beings. New England Journal of Medicine, 303(6), 322–324. https://doi.org/10.1056/nejm198008073030607
- Newman, D., Sauve, M. J., Herre, J., Langberg, J. J., Lee, M. A., Titus, C., Franklin, J., Scheinman, M. M., & Griffin, J. C. (1992). Survival after implantation of the cardioverter defibrillator. The American Journal of Cardiology, 69(9), 899–903. https://doi.org/10.1016/0002-9149(92)90789-2
- Winkle, R. A., Mead, R. H., Ruder, M. A., Gaudiani, V. A., Smith, N. A., Buch, W. S., Schmidt, P., & Shipman, T. (1989). Long-term outcome with the automatic implantable cardioverter-defibrillator. Journal of the American College of Cardiology, 13(6), 1353–1361. https://doi.org/10.1016/0735-1097(89)90310-0
- Tachycardia detection and treatment with a third‐generation implantable cardioverter‐defibrillator: An international experience, (abstract)
Citations
- Markman, T. M., & Callahan, T. D. (2026). The extravascular ICD: Evolving paradigms in sudden death prevention. Heart Rhythm, 23(1), 186–193. https://doi.org/10.1016/j.hrthm.2025.09.038
- https://doi.org/10.1111/jce.15916
- https://doi.org/10.4330/wjc.v9.i5.429
- Janardhan, A. H., Li, W., Fedorov, V. V., Yeung, M., Wallendorf, M. J., Schuessler, R. B., & Efimov, I. R. (2012). A Novel Low-Energy Electrotherapy That Terminates Ventricular Tachycardia With Lower Energy Than a Biphasic Shock When Antitachycardia Pacing Fails. Journal of the American College of Cardiology, 60(23), 2393–2398. https://doi.org/10.1016/j.jacc.2012.08.1001
- Schoenhard, J. A., & Zei, P. C. (2011). Reducing ICD Shocks for Ventricular Arrhythmias. Cardiac Electrophysiology Clinics, 3(3), 493–502. https://doi.org/10.1016/j.ccep.2011.05.013
- https://doi.org/10.1111/j.1540-8159.2010.03004.x
- https://doi.org/10.1111/j.1540-8159.2008.01147.x
- Sauer, W. H., & Callans, D. J. (2007). The Implantable Cardioverter-Defibrillator. In Cardiovascular Medicine (pp. 2119–2138). Springer London. https://doi.org/10.1007/978-1-84628-715-2_104
- Wathen, M. (2007). Implantable cardioverter defibrillator shock reduction using new antitachycardia pacing therapies. American Heart Journal, 153(4), 44–52. https://doi.org/10.1016/j.ahj.2007.01.020
- https://doi.org/10.1152/ajpheart.01300.2005
- https://doi.org/10.1161/01.cir.0000145610.64014.e4
- https://doi.org/10.1067/mtc.2002.121160
- https://doi.org/10.1016/s0033-0620(98)80058-3
- https://doi.org/10.1111/j.1540-8167.1997.tb00822.x
- https://doi.org/10.1016/s0002-9149(97)89311-3
- https://doi.org/10.1111/j.1540-8159.1995.tb02478.x
- https://doi.org/10.1136/hrt.73.2.158
- Misaki, T., Watanabe, G., Iwa, T., Tsubota, M., Ohtake, H., Yamamoto, K., & Watanabe, Y. (1994). Surgical treatment of arrhythmogenic right ventricular dysplasia: Long-term outcome. The Annals of Thoracic Surgery, 58(5), 1380–1385. https://doi.org/10.1016/0003-4975(94)91918-6
- https://doi.org/10.1016/s0899-5885(18)30510-0
- Porterfield, J. G., Porterfield, L. M., Smith, B. A., & Bray, L. (1993). Experience with three different third-generation cardioverter-defibrillators in patients with coronary artery disease or cardiomyopathy. The American Journal of Cardiology, 72(3), 301–304. https://doi.org/10.1016/0002-9149(93)90676-4