Published September 3, 2020 | Journal article | Version v1 | https://doi.org/10.1055/a-1238-2802

Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography – Evaluation Using an ACR Phantom

Description

Purpose To evaluate the influence of patient alignment and thereby heel effect on the image quality (IQ) of C-arm flat-panel detector computed tomography (CACT). Materials and Methods An ACR phantom placed in opposite directions along the z-axis (setup A and B) on the patient support was imaged using CACT. Image acquisition was performed with three different image acquisition protocols. The images were reconstructed with four convolution kernels. IQ was assessed in terms of high contrast using the modulation transfer function (MTF) and low contrast by assessing the image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratios (CNR) as well as the reliability of density measurements. Furthermore, the dose intensity profiles were measured free-in-air. Results The MTF in setup B is higher than the MTF measured in setup A (p < 0.01). The image noises measured in setup A for the air and bone inserts were higher compared to those measured in setup B (p > 0.05). Opposite behavior has been observed for the polyethylene, water-equivalent and acrylic inserts. The SNR for all inserts is inversely related to the image noise. A systematically increasing or decreasing trend of CNR could not be observed (p > 0.05). The intensity profile measured by the detector system free-in-air showed that the anode heel effect is perpendicular to the z-axis. Conclusion The patient alignment has a minor influence on the IQ of CACT. This effect is not based on the X-ray anode heel effect but is caused mainly by the non-symmetrical rotation of CACT. Key Points:  Citation Format

Additional details

Dates

Issued
September 3, 2020

References

  1. Smyth, J., Sutton, D., & Houston, J. (2006). Evaluation of the quality of CT-like images obtained using a commercial flat panel detector system. Biomedical Imaging and Intervention Journal, 2(4). https://doi.org/10.2349/biij.2.4.e48
  2. Rotational flat-panel computed tomography in diagnostic and interventional neuroradiology
  3. Braun, H., Kyriakou, Y., Kachelrieß, M., & Kalender, W. A. (2010). The influence of the heel effect in cone-beam computed tomography: artifacts in standard and novel geometries and their correction. Physics in Medicine and Biology, 55(19), 6005–6021. https://doi.org/10.1088/0031-9155/55/19/024
  4. Ott, S., Struffert, T., Saake, M., Gölitz, P., Adamek, E., & Doerfler, A. (2015). Influence of different reconstruction parameters in the visualization of intracranial stents using C-arm flat panel CT angiography: experience in an animal model. Acta Radiologica, 57(2), 233–240. https://doi.org/10.1177/0284185115571988
  5. Werncke, T., Sonnow, L., Meyer, B. C., Lüpke, M., Hinrichs, J., Wacker, F. K., & von Falck, C. (2017). Ultra-high resolution C-Arm CT arthrography of the wrist: Radiation dose and image quality compared to conventional multidetector computed tomography. European Journal of Radiology, 89, 191–199. https://doi.org/10.1016/j.ejrad.2017.02.009
  6. Jones, A. K., & Mahvash, A. (2012). Evaluation of the potential utility of flat panel CT for quantifying relative contrast enhancement. Medical Physics, 39(7Part1), 4149–4154. https://doi.org/10.1118/1.4725711
  7. Orth, R. C., Wallace, M. J., & Kuo, M. D. (2008). C-arm Cone-beam CT: General Principles and Technical Considerations for Use in Interventional Radiology. Journal of Vascular and Interventional Radiology, 19(6), 814–820. https://doi.org/10.1016/j.jvir.2008.02.002
  8. Fahrig, R., Dixon, R., Payne, T., Morin, R. L., Ganguly, A., & Strobel, N. (2006). Dose and image quality for a cone‐beam C‐arm CT system. Medical Physics, 33(12), 4541–4550. https://doi.org/10.1118/1.2370508
  9. Mittal, B., Rana, N., Rawat, D., Parmar, M., Dhawan, D., & Bhati, A. (2015). Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography. Journal of Medical Physics, 40(4), 198. https://doi.org/10.4103/0971-6203.170790
  10. Zhang, G., Marshall, N., Jacobs, R., Liu, Q., & Bosmans, H. (2013). Bowtie filtration for dedicated cone beam CT of the head and neck: a simulation study. The British Journal of Radiology, 86(1028), 20130002. https://doi.org/10.1259/bjr.20130002
  11. Mori, S., Endo, M., Nishizawa, K., Ohno, M., Miyazaki, H., Tsujita, K., & Saito, Y. (2005). Prototype heel effect compensation filter for cone-beam CT. Physics in Medicine and Biology, 50(22), N359–N370. https://doi.org/10.1088/0031-9155/50/22/n02
  12. Kusano, Y., Uesaka, S., Yajima, K., Kumagai, M., Mizuno, H., & Mori, S. (2012). Positional dependence of the CT number with use of a cone-beam CT scanner for an electron density phantom in particle beam therapy. Radiological Physics and Technology, 6(1), 241–247. https://doi.org/10.1007/s12194-012-0194-x
  13. McCollough, C. H., Bruesewitz, M. R., McNitt‐Gray, M. F., Bush, K., Ruckdeschel, T., Payne, J. T., Brink, J. A., & Zeman, R. K. (2004). The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoid. Medical Physics, 31(9), 2423–2442. https://doi.org/10.1118/1.1769632
  14. Hinrichs, J. B., von Falck, C., Hoeper, M. M., Olsson, K. M., Wacker, F. K., Meyer, B. C., & Renne, J. (2016). Pulmonary Artery Imaging in Patients with Chronic Thromboembolic Pulmonary Hypertension: Comparison of Cone-Beam CT and 64-Row Multidetector CT. Journal of Vascular and Interventional Radiology, 27(3), 361–368.e2. https://doi.org/10.1016/j.jvir.2015.11.046
  15. Dixon, R. L., Munley, M. T., & Bayram, E. (2005). An improved analytical model for CT dose simulation with a new look at the theory of CT dose. Medical Physics, 32(12), 3712–3728. https://doi.org/10.1118/1.2122507
  16. Dixon, R. L., & Boone, J. M. (2011). Analytical equations for CT dose profiles derived using a scatter kernel of Monte Carlo parentage with broad applicability to CT dosimetry problems. Medical Physics, 38(7), 4251–4264. https://doi.org/10.1118/1.3600518
  17. Droege, R. T., & Morin, R. L. (1982). A practical method to measure the MTF of CT scanners. Medical Physics, 9(5), 758–760. https://doi.org/10.1118/1.595124
  18. Grimmer, R., & Kachelrieß, M. (2011). Empirical binary tomography calibration (EBTC) for the precorrection of beam hardening and scatter for flat panel CT. Medical Physics, 38(4), 2233–2240. https://doi.org/10.1118/1.3561506
  19. Hunter, A., & McDavid, W. (2012). Characterization and correction of cupping effect artefacts in cone beam CT. Dentomaxillofacial Radiology, 41(3), 217–223. https://doi.org/10.1259/dmfr/19015946
  20. Dwivedi, N., Nagarajappa, A., & Tiwari, R. (2015). Artifacts: The downturn of CBCT image. Journal of International Society of Preventive and Community Dentistry, 5(6), 440. https://doi.org/10.4103/2231-0762.170523
  21. Barrett, J. F., & Keat, N. (2004). Artifacts in CT: Recognition and Avoidance. RadioGraphics, 24(6), 1679–1691. https://doi.org/10.1148/rg.246045065
  22. Bai, M., Liu, B., Mu, H., Liu, X., & Jiang, Y. (2012). The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): A phantom study. European Journal of Radiology, 81(11), 3577–3580. https://doi.org/10.1016/j.ejrad.2011.09.006
  23. Schegerer, A. A., Lechel, U., Ritter, M., Weisser, G., Fink, C., & Brix, G. (2014). Dose and Image Quality of Cone-Beam Computed Tomography as Compared With Conventional Multislice Computed Tomography in Abdominal Imaging. Investigative Radiology, 49(10), 675–684. https://doi.org/10.1097/rli.0000000000000069
  24. Jones, A. K., & Odisio, B. C. (2020). Comparison of radiation dose and image quality between flat panel computed tomography and multidetector computed tomography in a hybrid CT‐angiography suite. Journal of Applied Clinical Medical Physics, 21(2), 121–127. https://doi.org/10.1002/acm2.12808