Development and performance test of an online blood sampling system for determination of the arterial input function in rats
Authors/Creators
Description
Abstract Background For positron emission tomography (PET) kinetic modelling, an accurate determination of the arterial input function is required. In this study, a blood sampling system was developed and tested using different radiotracers in rats. Methods The detector consists of pairs of lutetium yttrium oxyorthosilicate (LYSO) detectors, photomultiplier tubes and lead shield assembled within a steel casing working in coincidence mode. Rats were cannulated with microtubes in the femoral artery and vein for arterial blood sampling as well as administration of the PET tracers. Connected PTFE microtubes were centred between the LYSO crystals using a special holder. To enhance sensitivity, three layers with two coils were used. A flexible tube pump was used to ensure a constant blood flow. Performance of the detector was assessed with [18F]fludeoxyglucose (FDG), [18F]ciprofloxacin, (R)-[11C]verapamil, [11C]tariquidar, [11C]mephobarbital and [11C]MC113. Obtained input function curves were compared with manual samples drawn every 5 s during the first 3 min and further on at 5, 10, 20, 30, 40, 50 and 60 min after radiotracer injection. After manual sampling, an arterio/venous shunt was established. Shape and area-under-the-curve (AUC; Bq/μl*h) of the input functions were evaluated. Results The developed detector system provided an absolute sensitivity of 6.5%. Maximum peak values agreed well between manual samples and the detector with a mean difference of −0.4% ± 7.0% (max 12.0%, min −9.9%). AUC values also exhibited an excellent correlation (R = 0.996) between manual sampling and detector measurements with a mean difference of 9.3% ± 9.7% (max 24.1%, min −3.2%). The system was able to measure peak blood activity concentration levels of 110 to 2,000 Bq/μl which corresponds to injected activities from 5.5 to 100 MBq depending on the used radiotracer, applied volume and weight of the animal. Conclusions This study demonstrates that the developed blood sampling system can be used for in vivo small animal PET studies in rats in a reliable way. The usage of the systems enhances the accuracy of the input curve as handling of small blood samples especially with low activity (as for C-11) is prone to measurement errors. Additionally, the radiation dose of the experimenters can be reduced, as it is not required anymore to continuously draw samples where the personal is in close contact to the radioactive animals and blood.
Files
s40658-014-0106-8.pdf
Files
(3.9 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:f944e8ae699e010064defd75ebaed398
|
3.9 MB | Preview Download |
Additional details
Identifiers
Funding
Dates
- Issued
-
January 14, 2015
References
- Dupont, P., & Warwick, J. (2009). Kinetic modelling in small animal imaging with PET. Methods, 48(2), 98–103. https://doi.org/10.1016/j.ymeth.2009.03.008
- Walsh GM, Ferrone RA, Tsuchiya M, Woods EF, Deland EC. Hemodynamic and metabolic responses to repeated blood sampling in the rat. Am J Physiol. 1980;239:H805–9.
- McGuill, M. W., & Rowan, A. N. (1989). Biological Effects of Blood Loss: Implications for Sampling Volumes and Techniques * Commentary: H. Richard Adams. ILAR Journal, 31(4), 5–20. https://doi.org/10.1093/ilar.31.4.5
- Kim J, Herrero P, Sharp T, Laforest R, Rowland DJ, Tai YC, et al. Minimally invasive method of determining blood input function from PET images in rodents. J Nucl Med. 2006;47:330–6.
- Ferl, G. Z., Zhang, X., Wu, H.-M., & Huang, S.-C. (2007). Estimation of the 18F-FDG Input Function in Mice by Use of Dynamic Small-Animal PET and Minimal Blood Sample Data. Journal of Nuclear Medicine, 48(12), 2037–2045. https://doi.org/10.2967/jnumed.107.041061
- Eberl S, Anayat AR, Fulton RR, Hooper PK, Fulham MJ. Evaluation of two population-based input functions for quantitative neurological FDG PET studies. Eur J Nucl Med. 1997;24:299–304.
- Litton, J. E., & Eriksson, L. (1990). Transcutaneous measurement of the arterial input function in positron emission tomography. IEEE Transactions on Nuclear Science, 37(2), 627–628. https://doi.org/10.1109/23.106688
- Maramraju, S., Stoll, S., Woody, C., Schlyer, D., Schiffer, W., Lee, D., Dewey, S., & Vaska, P. (2006). A LSO Beta Microprobe for Measuring Input Functions for Quantitative Small Animal PET. 2006 IEEE Nuclear Science Symposium Conference Record, 2713–2716. https://doi.org/10.1109/nssmic.2006.356440
- Pain F, Laniece P, Mastrippolito R, Gervais P, Hantraye P, Besret L. Arterial input function measurement without blood sampling using a beta-microprobe in rats. J Nucl Med. 2004;45:1577–82.
- Hutchins, G. D., Hichwa, R. D., & Koeppe, R. A. (1986). A Continuous Flow Input Function Detector for H2 15O Blood Flow Studies in Positron Emission Tomography. IEEE Transactions on Nuclear Science, 33(1), 546–549. https://doi.org/10.1109/tns.1986.4337162
- Kanno, I., Iida, H., Miura, S., Murakami, M., Takahashi, K., Sasaki, H., Inugami, A., Shishido, F., & Uemura, K. (1987). A System for Cerebral Blood Flow Measurement Using an H 2 15 O Autoradiographic Method and Positron Emission Tomography. Journal of Cerebral Blood Flow & Metabolism, 7(2), 143–153. https://doi.org/10.1038/jcbfm.1987.37
- Nelson AD, Muzic RF, Miraldi F, Muswick GJ, Leisure GP, Voelker W. Continuous arterial positron monitor for quantitation in PET imaging. Am J Physiol Imaging. 1990;5:84–8.
- Ranicar AS, Williams CW, Schnorr L, Clark JC, Rhodes CG, Bloomfield PM, et al. The on-line monitoring of continuously withdrawn arterial blood during PET studies using a single BGO/photomultiplier assembly and non-stick tubing. Med Prog Technol. 1991;17:259–64.
- Eriksson, L., Holte, S., Bohm, C., Kesselberg, M., & Hovander, B. (1988). Automated blood sampling systems for positron emission tomography. IEEE Transactions on Nuclear Science, 35(1), 703–707. https://doi.org/10.1109/23.12815
- Eriksson, L., Ingvar, M., Rosenqvist, G., Stone-Elander, S., Ekdahl, T., & Kappel, P. (1995). Characteristics of a new automated blood sampling system for positron emission tomography. IEEE Transactions on Nuclear Science, 42(4), 1007–1011. https://doi.org/10.1109/23.467757
- Votaw JR, Shulman SD. Performance evaluation of the Pico-Count flow-through detector for use in cerebral blood flow PET studies. J Nucl Med. 1998;39:509–15.
- Tadokoro H, Yoshida K, Takami A, Shimada K, Kitsukawa S, Suzuki K, et al. A high detection efficiency BGO, flow-through, coincidence detection system for arterial blood sampling with PET: application and dispersion correction for small animal PET studies. In: Nuclear Science Symposium, 1997 IEEE; 9–15 Nov 1997. 1997. p. 1528–32 vol.1522.
- Ingvar, M., Eriksson, L., Rogers, G. A., Stone-Elander, S., & Widén, L. (1991). Rapid Feasibility Studies of Tracers for Positron Emission Tomography: High-Resolution PET in Small Animals with Kinetic Analysis. Journal of Cerebral Blood Flow & Metabolism, 11(6), 926–931. https://doi.org/10.1038/jcbfm.1991.157
- Yamamoto, S., Imaizumi, M., Shimosegawa, E., Kanai, Y., Sakamoto, Y., Minato, K., Shimizu, K., Senda, M., & Hatazawa, J. (2010). A compact and high sensitivity positron detector using dual-layer thin GSO scintillators for a small animal PET blood sampling system. Physics in Medicine and Biology, 55(13), 3813–3826. https://doi.org/10.1088/0031-9155/55/13/016
- Roehrbacher F. Development of an online flow-through system to determine the arterial input function for compartment modeling. PhD thesis. Vienna University of Technology, Vienna, Austria: Faculty of Physics; 2012.
- Knoll GF. Radiation detection and measurement. 3rd edition edn. University of Michigan, Ann Arbor, USA: John Wiley & Sons, Inc.; 2000.
- Munk, O. L., Keiding, S., & Bass, L. (2008). A method to estimate dispersion in sampling catheters and to calculate dispersion-free blood time-activity curves: A method to calculate dispersion-free blood time-activity curves. Medical Physics, 35(8), 3471–3481. https://doi.org/10.1118/1.2948391
- Convert, L., Morin-Brassard, G., Cadorette, J., Rouleau, D., Croteau, E., Archambault, M., Fontaine, R., & Lecomte, R. (2007). A Microvolumetric $\beta$ Blood Counter for Pharmacokinetic PET Studies in Small Animals. IEEE Transactions on Nuclear Science, 54(1), 173–180. https://doi.org/10.1109/tns.2006.889166
- Hamacher K, Coenen HH, Stocklin G. Efficient stereospecific synthesis of no-carrier-added 2-[18F]-fluoro-2-deoxy-D-glucose using aminopolyether supported nucleophilic substitution. J Nucl Med. 1986;27:235–8.
- BRUNNER, M., LANGER, O., SUNDERPLASSMANN, R., DOBROZEMSKY, G., MULLER, U., WADSAK, W., KRCAL, A., KARCH, R., MANNHALTER, C., & DUDCZAK, R. (2005). Influence of functional haplotypes in the drug transporter gene on central nervous system drug distribution in humans. Clinical Pharmacology & Therapeutics, 78(2), 182–190. https://doi.org/10.1016/j.clpt.2005.04.011
1–25 of 38
Citations
- https://doi.org/10.1186/s40658-025-00789-6
- Yusoff, R., Baco, S., Rahman, A. B. A., Ghazali, K., Gabda, D., & Al-Somali, F. (2023). Design study on the microfluidic scintillator detector using GEANT4. AIP Conference Proceedings, 2848, 020005. https://doi.org/10.1063/5.0127952
- Convert, L., Sarrhini, O., Paillé, M., Salem, N., Charette, P. G., & Lecomte, R. (2022). The ultra high sensitivity blood counter: a compact, MRI-compatible, radioactivity counter for pharmacokinetic studies in μl volumes. Biomedical Physics & Engineering Express, 8(3), 035022. https://doi.org/10.1088/2057-1976/ac4c29
- Serrano-Sosa, M., & Huang, C. (2021). Motion Correction in PET/MRI. In Hybrid PET/MR Neuroimaging (pp. 27–34). Springer International Publishing. https://doi.org/10.1007/978-3-030-82367-2_4
- Rey-Bretal, D., Moscoso, A., Gómez-Lado, N., Fernández-Ferreiro, A., Silva-Rodríguez, J., Ruibal, Á., & Aguiar, P. (2020). Feasibility of Longitudinal Brain PET with Real-Time Arterial Input Function in Rats. Molecular Imaging and Biology, 23(3), 350–360. https://doi.org/10.1007/s11307-020-01556-y
- https://doi.org/10.1186/s40658-020-0277-4
- Hou, X., Zhao, W., Beauregard, J.-M., & Celler, A. (2019). Personalized kidney dosimetry in 177 Lu-octreotate treatment of neuroendocrine tumours: a comparison of kidney dosimetry estimates based on a whole organ and small volume segmentations. Physics in Medicine & Biology, 64(17), 175004. https://doi.org/10.1088/1361-6560/ab32a1
- https://doi.org/10.1186/s40658-019-0263-x
- Goenka, A. H., Nanni, C., & Venkatesh, S. K. (2018). Rectal Cancer. In PET/MRI in Oncology (pp. 305–319). Springer International Publishing. https://doi.org/10.1007/978-3-319-68517-5_17
- https://doi.org/10.2967/jnumed.117.204164
- Öktem, O., Chen, C., Onur Domaniç, N., Ravikumar, P., & Bajaj, C. (2017). Shape-based image reconstruction using linearized deformations. Inverse Problems, 33(3), 035004. https://doi.org/10.1088/1361-6420/aa55af
- https://doi.org/10.1186/s40658-017-0184-5
- https://doi.org/10.2967/jnumed.115.162768