The Hannover Coupler V2: Audiological outcomes of a round window coupler for the Vibrant Soundbridge
Authors/Creators
-
1.
Medizinische Hochschule Hannover
- 2. Department of Otorhinolaryngology German Hearing Center and Institute of Audioneurotechnology (VIANNA), Hannover Medical School Hannover Germany
-
3.
Hearing4all
Description
AbstractIntroductionThe Hannover Coupler version 2 (HC2) was designed to (1) adapt the coupler geometry to the round window (RW) niche (2) to stabilize the floating mass transducer, and (3) to control static coupling forces to the RW. First audiological outcomes with a custom‐made HC2 are reported here.Material and MethodsTen patients were enrolled in our site‐initiated, prospective study. To assess audiological outcomes up to 6 months, preoperative and postoperative hearing thresholds, word recognition score (WRS) at 65 dB SPL and the speech recognition threshold in quiet and noise were performed. The effective gain (EG) and the coupling efficiency were calculated.ResultsOne revision surgery had to be performed during the study period and a significant, but clinically not relevant bone conduction thresholds change was observed at 4 and 6 kHz at 6‐month follow‐up. At 6 months, the median WRS (n = 10) improved significantly from 0% to 80%. The median speech reception threshold in noise improved significantly from 11.6 to −2.4 dB SNR, and in quiet significantly from 79.6 to 44.4 dB SPL. The average EG of −1.3 dB indicated a closure of the air bone gap. The determined average coupling efficiency of 23.3 dB was within the acceptance range suggested by the manufacturer.ConclusionFor patients with mixed hearing loss and multiple ear surgeries, the HC2 provided good and stable speech recognition results exceeding published results of RW coupling without a coupler or coupling with the RW soft coupler.
Additional details
Identifiers
- OpenAlex
- W4281619919
- PMID
- 35734043
- URL
- https://onlinelibrary.wiley.com/doi/10.1002/lio2.812
Funding
-
Deutsche Forschungsgemeinschaft
- 2177/1
-
Deutsche Forschungsgemeinschaft
- EXC2177/1
-
Deutsche Forschungsgemeinschaft
- EXC 2177: Hearing4all 390895286
-
Deutsche Forschungsgemeinschaft
- EXC2177
Dates
- Issued
-
May 28, 2022
References
- Ball, G. R. (2010). The Vibrant Soundbridge: Design and Development. In Advances in Oto-Rhino-Laryngology (pp. 1–13). S. Karger AG. https://doi.org/10.1159/000318516
- Ball, G., & Rose-Eichberger, K. (2021). DESIGN AND DEVELOPMENT OF THE VIBRANT SOUNDBRIDGE – A 25-YEAR PERSPECTIVE. Journal of Hearing Science, 11(1), 9–20. https://doi.org/10.17430/jhs.2021.11.1.1
- https://doi.org/10.1055/s%E2%80%902007%E2%80%90996970 https://doi.org/10.1055/s%E2%80%902007%E2%80%90996970
- Colletti, V., Soli, S. D., Carner, M., & Colletti, L. (2006). Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. International Journal of Audiology, 45(10), 600–608. https://doi.org/10.1080/14992020600840903
- Dumon, T. (2007). Vibrant Soundbridge Middle Ear Implant in Otosclerosis. In Advances in Oto-Rhino-Laryngology (pp. 320–322). KARGER. https://doi.org/10.1159/000098852
- Cremers, C. W. R. J., Verhaegen, V. J. O., & Snik, A. F. M. (2009). The Floating Mass Transducer of the Vibrant Soundbridge Interposed Between the Stapes and Tympanic Membrane After Incus Necrosis. Otology & Neurotology, 30(1), 76–78. https://doi.org/10.1097/mao.0b013e31818f5790
- Hüttenbrink, K. B., Beutner, D., & Zahnert, T. (2010). Clinical Results with an Active Middle Ear Implant in the Oval Window. In Advances in Oto-Rhino-Laryngology (pp. 27–31). S. Karger AG. https://doi.org/10.1159/000318519
- Zehlicke, T., Dahl, R., Just, T., & Pau, H.-W. (2010). Vibroplasty involving direct coupling of the floating mass transducer to the oval window niche. The Journal of Laryngology & Otology, 124(7), 716–719. https://doi.org/10.1017/s0022215110000526
- Schwab, B., Salcher, R. B., Maier, H., & Kontorinis, G. (2012). Oval Window Membrane Vibroplasty for Direct Acoustic Cochlear Stimulation. Otology & Neurotology, 33(5), 804–809. https://doi.org/10.1097/mao.0b013e3182595471
- Beltrame, A. M., Martini, A., Prosser, S., Giarbini, N., & Streitberger, C. (2009). Coupling the Vibrant Soundbridge to Cochlea Round Window. Otology & Neurotology, 30(2), 194–201. https://doi.org/10.1097/mao.0b013e318180a495
- Zahnert, T., Löwenheim, H., Beutner, D., Hagen, R., Ernst, A., Pau, H.-W., Zehlicke, T., Kühne, H., Friese, N., Tropitzsch, A., Lüers, J.-C., Mlynski, R., Todt, I., & Hüttenbrink, K.-B. (2016). Multicenter Clinical Trial of Vibroplasty Couplers to Treat Mixed/Conductive Hearing Loss: First Results. Audiology and Neurotology, 21(4), 212–222. https://doi.org/10.1159/000444616
- https://doi.org/10.1007/s00405%E2%80%90020%E2%80%9006064%E2%80%90x https://doi.org/10.1007/s00405%E2%80%90020%E2%80%9006064%E2%80%90x
- Atturo, F., Barbara, M., & Rask-Andersen, H. (2014). Is the Human Round Window Really Round? An Anatomic Study With Surgical Implications. Otology & Neurotology, 35(8), 1354–1360. https://doi.org/10.1097/mao.0000000000000332
- Mancheño, M., Aristegui, M., & Sañudo, J. R. (2017). Round and Oval Window Anatomic Variability: Its Implication for the Vibroplasty Technique. Otology & Neurotology, 38(5), e50–e57. https://doi.org/10.1097/mao.0000000000001385
- Pau, H. W., Just, T., Bornitz, M., Lasurashvilli, N., & Zahnert, T. (2007). Noise Exposure of the Inner Ear During Drilling a Cochleostomy for Cochlear Implantation. The Laryngoscope, 117(3), 535–540. https://doi.org/10.1097/mlg.0b013e31802f4169
- The active middle ear implant for the rehabilitation of sensorineural, mixed and conductive hearing losses
- https://doi.org/10.1007/s00405%E2%80%90013%E2%80%902432%E2%80%901 https://doi.org/10.1007/s00405%E2%80%90013%E2%80%902432%E2%80%901
- Olszewski, L., Jedrzejczak, W. W., Piotrowska, A., & Skarzynski, H. (2017). Round window stimulation with the Vibrant Soundbridge: Comparison of direct and indirect coupling. The Laryngoscope, 127(12), 2843–2849. https://doi.org/10.1002/lary.26536
- Schwab, B., Grigoleit, S., & Teschner, M. (2013). Do We Really Need a Coupler for the Round Window Application of an AMEI?. Otology & Neurotology, 34(7), 1181–1185. https://doi.org/10.1097/mao.0b013e31829b57c2
- https://doi.org/10.1007/s00405%E2%80%90016%E2%80%903997%E2%80%902 https://doi.org/10.1007/s00405%E2%80%90016%E2%80%903997%E2%80%902
- Maier, H., Salcher, R., Schwab, B., & Lenarz, T. (2013). The effect of static force on round window stimulation with the direct acoustic cochlea stimulator. Hearing Research, 301, 115–124. https://doi.org/10.1016/j.heares.2012.12.010
- Salcher, R., Schwab, B., Lenarz, T., & Maier, H. (2014). Round window stimulation with the floating mass transducer at constant pretension. Hearing Research, 314, 1–9. https://doi.org/10.1016/j.heares.2014.04.001
- Gostian, A. O., Pazen, D., Ortmann, M., Anagiotos, A., Schwarz, D., Hüttenbrink, K. B., & Beutner, D. (2016). Loads and Coupling Modalities Influence the Performance of the Floating Mass Transducer as a Round Window Driver. Otology & Neurotology, 37(5), 524–532. https://doi.org/10.1097/mao.0000000000001028
- Müller, M., Salcher, R., Lenarz, T., & Maier, H. (2017). The Hannover Coupler: Controlled Static Prestress in Round Window Stimulation With the Floating Mass Transducer. Otology & Neurotology, 38(8), 1186–1192. https://doi.org/10.1097/mao.0000000000001484
- Tringali, S., Koka, K., Deveze, A., Holland, N. J., Jenkins, H. A., & Tollin, D. J. (2010). Round Window Membrane Implantation with an Active Middle Ear Implant: A Study of the Effects on the Performance of Round Window Exposure and Transducer Tip Diameter in Human Cadaveric Temporal Bones. Audiology and Neurotology, 15(5), 291–302. https://doi.org/10.1159/000283006
1–25 of 45
Citations
- Liu, H., Ghoncheh, M., Schurzig, D., Lenarz, T., Prenzler, N., & Maier, H. (2026). Effect of coupler tip size on round window stimulation with the floating mass transducer under controlled preload. Hearing Research, 473, Article 109565. https://doi.org/10.1016/j.heares.2026.109565
- Prediger, M. S., Müller, E., Glukhovskoy, A., Bethmann, K., Berta, A. B., Wurz, M. C., & Maier, H. (2025). Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load. Biomedical Microdevices, 28(1), Article 1. https://doi.org/10.1007/s10544-025-00766-x
- Liu, H., Liu, Z., Liu, J., Lenarz, T., & Maier, H. (2025). Effect of electromagnetic transducer design parameters on round-window stimulation in otosclerosis: a nonlinear dynamic analysis. Nonlinear Dynamics, 113(14), 18727–18744. https://doi.org/10.1007/s11071-025-11149-5
- Knölke, N., Busch, S., Lenarz, T., & Maier, H. (2025). Vibrant Soundbridge Round Window Vibroplasty: Safety, Coupling Efficiency, and Speech Outcome of the Most Common Coupling Modalities. Audiology and Neurotology, 30(5), 392–403. https://doi.org/10.1159/000545335
- Maier, H., Lenarz, T., & Busch, S. (2024). An easy method to determine crucial AMEI performance parameters from clinical routine data in individuals - Part 1: maximum output. International Journal of Audiology, 64(3), 290–297. https://doi.org/10.1080/14992027.2024.2341100
- Liu, H., Liu, Z., Liu, J., Lenarz, T., & Maier, H. (2024). Effect of electromagnetic transducer design parameters on round-window stimulation in otosclerosis: A nonlinear dynamic analysis. In Research Square. Springer Science and Business Media LLC. https://doi.org/10.21203/rs.3.rs-4956003/v1