A realistic phantom of the human head for PET-MRI
Description
Abstract Background The combination of positron emission tomography (PET) and magnetic resonance imaging (MRI) (PET-MRI) is a unique hybrid imaging modality mainly used in oncology and neurology. The MRI-based attenuation correction (MRAC) is crucial for correct quantification of PET data. A suitable phantom to validate quantitative results in PET-MRI is currently missing. In particular, the correction of attenuation due to bone is usually not verified by commonly available phantoms. The aim of this work was, thus, the development of such a phantom and to explore whether such a phantom might be used to validate MRACs. Method Various materials were investigated for their attenuation and MR properties. For the substitution of bone, water-saturated gypsum plaster was used. The attenuation of 511 keV annihilation photons was regulated by addition of iodine. Adipose tissue was imitated by silicone and brain tissue by agarose gel, respectively. The practicability with respect to the comparison of MRACs was checked as follows: A small flask inserted into the phantom and a large spherical phantom (serving as a reference with negligible error in MRAC) were filled with the very same activity concentration. The activity concentration was measured and compared using clinical protocols on PET-MRI and different built-in and offline MRACs. The same measurements were carried out using PET-CT for comparison. Results The phantom imitates the human head in sufficient detail. All tissue types including bone were detected as such so that the phantom-based comparison of the quantification accuracy of PET-MRI was possible. Quantitatively, the activity concentration in the brain, which was determined using different MRACs, showed a deviation of about 5% on average and a maximum deviation of 11% compared to the spherical phantom. For PET-CT, the deviation was 5%. Conclusions The comparatively small error in quantification indicates that it is possible to construct a brain PET-MRI phantom that leads to MR-based attenuation-corrected images with reasonable accuracy.
Additional details
Identifiers
Dates
- Issued
-
August 5, 2020
References
- Rausch, I., Rischka, L., Ladefoged, C. N., Furtner, J., Fenchel, M., Hahn, A., Lanzenberger, R., Mayerhoefer, M. E., Traub-Weidinger, T., & Beyer, T. (2017). PET/MRI for Oncologic Brain Imaging: A Comparison of Standard MR-Based Attenuation Corrections with a Model-Based Approach for the Siemens mMR PET/MR System. Journal of Nuclear Medicine, 58(9), 1519–1525. https://doi.org/10.2967/jnumed.116.186148
- Andersen, F. L., Ladefoged, C. N., Beyer, T., Keller, S. H., Hansen, A. E., Højgaard, L., Kjær, A., Law, I., & Holm, S. (2014). Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring bone. NeuroImage, 84, 206–216. https://doi.org/10.1016/j.neuroimage.2013.08.042
- Hofmann, M., Steinke, F., Scheel, V., Charpiat, G., Farquhar, J., Aschoff, P., Brady, M., Schölkopf, B., & Pichler, B. J. (2008). MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration. Journal of Nuclear Medicine, 49(11), 1875–1883. https://doi.org/10.2967/jnumed.107.049353
- Hoffman, E. J., Cutler, P. D., Digby, W. M., & Mazziotta, J. C. (1990). 3-D phantom to simulate cerebral blood flow and metabolic images for PET. IEEE Transactions on Nuclear Science, 37(2), 616–620. https://doi.org/10.1109/23.106686
- Iida, H., Hori, Y., Ishida, K., Imabayashi, E., Matsuda, H., Takahashi, M., Maruno, H., Yamamoto, A., Koshino, K., Enmi, J., Iguchi, S., Moriguchi, T., Kawashima, H., & Zeniya, T. (2012). Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour. Annals of Nuclear Medicine, 27(1), 25–36. https://doi.org/10.1007/s12149-012-0655-7
- Sattler, B., Jochimsen, T., Barthel, H., Sommerfeld, K., Stumpp, P., Hoffmann, K.-T., Gutberlet, M., Villringer, A., Kahn, T., & Sabri, O. (2012). Physical and organizational provision for installation, regulatory requirements and implementation of a simultaneous hybrid PET/MR-imaging system in an integrated research and clinical setting. Magnetic Resonance Materials in Physics, Biology and Medicine, 26(1), 159–171. https://doi.org/10.1007/s10334-012-0347-2
- Teuho, J., Johansson, J., Linden, J., Saunavaara, V., Tolvanen, T., & Teräs, M. (2014). Specification and estimation of sources of bias affecting neurological studies in PET/MR with an anatomical brain phantom. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 734, 179–184. https://doi.org/10.1016/j.nima.2013.09.002
- Teuho, J., Johansson, J., Linden, J., Hansen, A. E., Holm, S., Keller, S. H., Delso, G., Veit-Haibach, P., Magota, K., Saunavaara, V., Tolvanen, T., Teräs, M., & Iida, H. (2016). Effect of Attenuation Correction on Regional Quantification Between PET/MR and PET/CT: A Multicenter Study Using a 3-Dimensional Brain Phantom. Journal of Nuclear Medicine, 57(5), 818–824. https://doi.org/10.2967/jnumed.115.166165
- Keereman, V., Mollet, P., Fierens, Y., Espana, S., & Vandenberghe, S. (2011). Design of a realistic PET-CT-MRI phantom. 2011 IEEE Nuclear Science Symposium Conference Record, 3173–3177. https://doi.org/10.1109/nssmic.2011.6153651
- Dixon, W. T. (1984). Simple proton spectroscopic imaging. Radiology, 153(1), 189–194. https://doi.org/10.1148/radiology.153.1.6089263
- Martinez-Möller, A., Souvatzoglou, M., Delso, G., Bundschuh, R. A., Chefd'hotel, C., Ziegler, S. I., Navab, N., Schwaiger, M., & Nekolla, S. G. (2009). Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT Data. Journal of Nuclear Medicine, 50(4), 520–526. https://doi.org/10.2967/jnumed.108.054726
- Holmes, J. E., & Bydder, G. M. (2005). MR imaging with ultrashort TE (UTE) pulse sequences: Basic principles. Radiography, 11(3), 163–174. https://doi.org/10.1016/j.radi.2004.07.007
- Reichert, I. L. H., Robson, M. D., Gatehouse, P. D., He, T., Chappell, K. E., Holmes, J., Girgis, S., & Bydder, G. M. (2005). Magnetic resonance imaging of cortical bone with ultrashort TE pulse sequences. Magnetic Resonance Imaging, 23(5), 611–618. https://doi.org/10.1016/j.mri.2005.02.017
- Delso, G., Kemp, B., Kaushik, S., Wiesinger, F., & Sekine, T. (2018). Improving PET/MR brain quantitation with template-enhanced ZTE. NeuroImage, 181, 403–413. https://doi.org/10.1016/j.neuroimage.2018.07.029
- Ladefoged, C. N., Benoit, D., Law, I., Holm, S., Kjær, A., Højgaard, L., Hansen, A. E., & Andersen, F. L. (2015). Region specific optimization of continuous linear attenuation coefficients based on UTE (RESOLUTE): application to PET/MR brain imaging. Physics in Medicine and Biology, 60(20), 8047–8065. https://doi.org/10.1088/0031-9155/60/20/8047
- Izquierdo-Garcia, D., Hansen, A. E., Förster, S., Benoit, D., Schachoff, S., Fürst, S., Chen, K. T., Chonde, D. B., & Catana, C. (2014). An SPM8-Based Approach for Attenuation Correction Combining Segmentation and Nonrigid Template Formation: Application to Simultaneous PET/MR Brain Imaging. Journal of Nuclear Medicine, 55(11), 1825–1830. https://doi.org/10.2967/jnumed.113.136341
- Burgos, N., Cardoso, M. J., Thielemans, K., Modat, M., Pedemonte, S., Dickson, J., Barnes, A., Ahmed, R., Mahoney, C. J., Schott, J. M., Duncan, J. S., Atkinson, D., Arridge, S. R., Hutton, B. F., & Ourselin, S. (2014). Attenuation Correction Synthesis for Hybrid PET-MR Scanners: Application to Brain Studies. IEEE Transactions on Medical Imaging, 33(12), 2332–2341. https://doi.org/10.1109/tmi.2014.2340135
- Chandramohan, D., Cao, P., Han, M., An, H., Sunderland, J. J., Kinahan, P. E., Laforest, R., Hope, T. A., & Larson, P. E. Z. (2020). Bone material analogues for PET/MRI phantoms. Medical Physics, 47(5), 2161–2170. https://doi.org/10.1002/mp.14079
- Yushkevich, P. A., Piven, J., Hazlett, H. C., Smith, R. G., Ho, S., Gee, J. C., & Gerig, G. (2006). User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. NeuroImage, 31(3), 1116–1128. https://doi.org/10.1016/j.neuroimage.2006.01.015
- Riehle, N., Götz, T., Kandelbauer, A., Tovar, G. E. M., & Lorenz, G. (2018). Data on the synthesis and mechanical characterization of polysiloxane-based urea-elastomers prepared from amino-terminated polydimethylsiloxanes and polydimethyl-methyl-phenyl-siloxane-copolymers. Data in Brief, 18, 1784–1794. https://doi.org/10.1016/j.dib.2018.04.083
- Reeder, S. B., Cruite, I., Hamilton, G., & Sirlin, C. B. (2011). Quantitative assessment of liver fat with magnetic resonance imaging and spectroscopy. Journal of Magnetic Resonance Imaging, 34(4), 729–749. https://doi.org/10.1002/jmri.22580
- Mitchell, M. D., Kundel, H. L., Axel, L., & Joseph, P. M. (1986). Agarose as a tissue equivalent phantom material for NMR imaging. Magnetic Resonance Imaging, 4(3), 263–266. https://doi.org/10.1016/0730-725x(86)91068-4
- Kraft, K. A., Fatouros, P. P., Clarke, G. D., & Kishore, P. R. S. (1987). An MRI phantom material for quantitative relaxometry. Magnetic Resonance in Medicine, 5(6), 555–562. https://doi.org/10.1002/mrm.1910050606
- Hellerbach, A., Schuster, V., Jansen, A., Sommer, J., & Oreja-Guevara, C. (2013). MRI Phantoms – Are There Alternatives to Agar?. PLoS ONE, 8(8), e70343. https://doi.org/10.1371/journal.pone.0070343
- McKee, B. T. A., Clack, R., Harvey, P. J., Hiltz, L. G., Hogan, M. J., & Howse, D. C. (1991). Accurate attenuation correction for a 3D PET system. Physics in Medicine and Biology, 36(5), 603–619. https://doi.org/10.1088/0031-9155/36/5/004
1–25 of 29
Citations
- https://doi.org/10.3390/tomography12040049
- Deng, L., Wu, T., Wu, F., Xiong, L., Yang, H., Chen, Q., & Liao, Y. (2025). Anthropomorphic Head MRI Phantoms: Technical Development, Brain Imaging Applications, and Future Prospects. Journal of Magnetic Resonance Imaging, 62(6), 1579–1602. https://doi.org/10.1002/jmri.29818
- https://doi.org/10.2478/ijssis-2024-0028
- Akamatsu, G., Tsutsui, Y., Daisaki, H., Mitsumoto, K., Baba, S., & Sasaki, M. (2023). A review of harmonization strategies for quantitative PET. Annals of Nuclear Medicine, 37(2), 71–88. https://doi.org/10.1007/s12149-022-01820-x
- https://doi.org/10.3389/fphy.2023.1134588
- Kilian, D., Kilian, W., Troia, A., Nguyen, T.-D., Ittermann, B., Zilberti, L., & Gelinsky, M. (2022). 3D Extrusion Printing of Biphasic Anthropomorphic Brain Phantoms Mimicking MR Relaxation Times Based on Alginate-Agarose-Carrageenan Blends. ACS Applied Materials & Interfaces, 14(43), 48397–48415. https://doi.org/10.1021/acsami.2c12872
- https://doi.org/10.2147/ijn.s362192
- https://doi.org/10.1186/s40658-021-00364-9
- https://doi.org/10.1186/s40658-021-00408-0
- https://doi.org/10.2967/jnumed.120.261881