Chronic Experiences with a Single Lead Dual Chamber Implantable Cardioverter Defibrillator System
Description
Monitoring of atrial rhythm in patients implanted with ICDs may improve accuracy in identifying supraventricular arrhythmias and, therefore, prevent inappropriate therapies. Since difficulties were found in dual chamber ICDs with separate leads, a new designed single lead dual chamber ICD system was tested. Twenty‐five patients implanted with a Deikos A+ (single coil defibrillation lead with two atrial sensing rings combined with a dual chamber ICD with a high amplifying atrial channel) were tested. Atrial and ventricular signals were analyzed during sinus rhythm (SR) and sinus tachycardias (STs), atrial flutter and AF, and VT or VF. Follow‐ups were performed after 1, 3, 6, 9, and 12 months after implantation. Analysis of EGM amplitudes of stored episodes revealed that atrial signals during atrial flutter (2.1 ± 0.51 mV) were comparable to those of ST (2.2 ± 0.5 mV). Atrial amplitudes during AF were significantly lower (0.81 ± 0.5 mV, P < 0.01). During VF atrial "sinus" signals (2 ± 0.8 mV) were stable. Ventricular parameters did not differ from a standard ICD lead; defibrillation threshold was 11.4 ± 4.5 J (16 patients). During intraoperative and prehospital discharge measurements, 97.1% of SR‐P waves and 99.2% of atrial flutter waves were detected correctly. In AF 91.11% of atrial signals were detected. Analysis of 505 stored episodes showed that 96.8% of ST and 100% of atrial flutter and 100% of AF episodes have been classified correctly and no underdetection of VT/VF was found. The first experiences with the new VDD‐ICD system show an increase of the specificity to detect ventricular tachycardias to a level comparable to dual chamber ICDs with two leads. The reliability of this system has to be proven in a prospective randomized study. (PACE 2003; 26:1937–1943)
Additional details
Identifiers
Dates
- Issued
-
October 2003
References
- Nanthakumar, K., Paquette, M., Newman, D., Curtis Deno, D., Malden, L., Gunderson, B., Gilkerson, J., Greene, M., Heng, D., & Dorian, P. (2000). Inappropriate therapy from atrial fibrillation and sinus tachycardia in automated implantable cardioverter defibrillators. American Heart Journal, 139(5), 797–803. https://doi.org/10.1016/s0002-8703(00)90010-3
- Swerdlow, C. D., Ahern, T., Chen, P.-S., Hwang, C., Gang, E., Mandel, W., Kass, R. M., & Peter, C. T. (1994). Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. Journal of the American College of Cardiology, 24(2), 416–424. https://doi.org/10.1016/0735-1097(94)90298-4
- Swerdlow, C. D., Chen, P.-S., Kass, R. M., Allard, J. R., & Peter, C. T. (1994). Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. Journal of the American College of Cardiology, 23(6), 1342–1355. https://doi.org/10.1016/0735-1097(94)90376-x
- HIGGINS, S. L., LEE, R. S., & KRAMER, R. L. (1995). Stability: An ICD Detection Criterion for Discriminating Atrial Fibrillation From Ventricular Tachycardia. Journal of Cardiovascular Electrophysiology, 6(12), 1081–1088. https://doi.org/10.1111/j.1540-8167.1995.tb00385.x
- Brugada, J. (1999). Is inappropriate therapy a resolved issue with current implantable cardioverter defibrillators?. The American Journal of Cardiology, 83(5), 40–44. https://doi.org/10.1016/s0002-9149(98)00967-9
- NEUZNER, J., PITSCHNER, H. F., & SCHLEPPER, M. (1995). Programmable VT Detection Enhancements in Implantable Cardioverter Defibrillator Therapy. Pacing and Clinical Electrophysiology, 18(3), 539–547. https://doi.org/10.1111/j.1540-8159.1995.tb02564.x
- KORTE, T., JUNG, W., WOLPERT, C., SPEHL, S., SCHUMACHER, B., ESMAILZADEH, B., & LÜDERITZ, B. (1998). A New Classification Algorithm for Discrimination of Ventricular from Supraventricular Tachycardia in a Dual Chamher Implantable Cardioverter Defibrillator. Journal of Cardiovascular Electrophysiology, 9(1), 70–73. https://doi.org/10.1111/j.1540-8167.1998.tb00868.x
- OSSWALD, S., CRON, T. A., STERNS, L., ALT, E., STOTTS, L., & COENEN, M. (1998). New ICD‐Technologies: First Clinical Experience with Dual‐Chamber Sensing for Differentiation of Supraventricular Tachyarrhythmias. Pacing and Clinical Electrophysiology, 21(1), 292–295. https://doi.org/10.1111/j.1540-8159.1998.tb01107.x
- Nair, M., Saoudi, N., Kroiss, D., & Letac, B. (1997). Automatic Arrhythmia Identification Using Analysis of the Atrioventricular Association. Circulation, 95(4), 967–973. https://doi.org/10.1161/01.cir.95.4.967
- Schwacke, H., Drögemüller, A., Siemon, G., Werling, C., Saggau, W., Senges, J., & Seidl, K. (1999). Komplikationen mit Sonden bei 340 Patienten mit einem implantierbaren Kardioverter/Defibrillator. Zeitschrift für Kardiologie, 88(8), 559–565. https://doi.org/10.1007/s003920050326
- CHAUHAN, A., GRACE, A. A., NEWELL, S. A., STONE, D. L., SHAPIRO, L. M., SCHOFIELD, P. M., & PETCH, M. C. (1994). Early Complications After Dual Chamber Versus Single Chamber Pacemaker Implantation. Pacing and Clinical Electrophysiology, 17(11), 2012–2015. https://doi.org/10.1111/j.1540-8159.1994.tb03791.x
- GRADAUS, R., BLOCK, M., DORSZEWSKI, A., SCHRIEVER, C., HAMMEL, D., SCHELD, H. H., BORGGREFE, M., BREITHARDT, G., & BÖCKER, D. (2001). Implantation of a Dual Chamber Pacing and Sensing Single Pass Defibrillation Lead. Pacing and Clinical Electrophysiology, 24(4), 416–423. https://doi.org/10.1046/j.1460-9592.2001.00416.x
- Niehaus, M., Thuermer, G., Thamasett, S., Pfeiffer, D., Hoeper, K., Korte, T., Goerler, H., Belke, R., & Tebbenjohanns, J. (2001). Nutzen einer VDD-Defibrillationselektrode zur Aufzeichnung von atrialen Elektrogrammen bei Vorhofflimmern/-flattern. Herz, 26(1), 40–48. https://doi.org/10.1007/pl00002005
- NIEHAUS, M., SCHUCHERT, A., THAMASETT, S., PFEIFFER, D., KORTE, T., PICHLMAIER, M., PANNING, B., BELKE, R., & TEBBENJOHANNS, J. (2001). Multicenter Experiences with a Single Lead Electrode for Dual Chamber ICD Systems. Pacing and Clinical Electrophysiology, 24(10), 1489–1493. https://doi.org/10.1046/j.1460-9592.2001.01489.x
- Rey, J. L., Tribouilloy, C., Elghelbazouri, F., & Otmani, A. (1998). Single-lead VDD pacing: Long-term experience with four different systems. American Heart Journal, 135(6), 1036–1039. https://doi.org/10.1016/s0002-8703(98)70069-9
- FAERESTRAND, S., & OHM, O. (1998). Atrial Synchronous Ventricular Pacing with a Single Lead: Reliability of Atrial Sensing During Physical Activities, and Long‐term Stability of Atrial Sensing. Pacing and Clinical Electrophysiology, 21(1), 271–276. https://doi.org/10.1111/j.1540-8159.1998.tb01103.x
- Hazday, M. S., Mendelson, D., & Brownlee, R. R. (1998). Acute Evaluation of a Preformed Single-pass VDD/DDD Pacing Lead. Journal of Interventional Cardiac Electrophysiology, 2(2), 171–173. https://doi.org/10.1023/a:1009755616342
- SUN, Z., STJERNVALL, J., LAINE, P., & TOIVGNEN, L. (1998). Extensive Variation in the Signal Amplitude of the Atrial Floating VDD Pacing Electrode. Pacing and Clinical Electrophysiology, 21(9), 1760–1765. https://doi.org/10.1111/j.1540-8159.1998.tb00276.x
- CHAMBERLAIN‐WEBBER, R., BARNES, E., PAPOUCHADO, M., & CRICK, J. P. (1998). Long‐term Survival of VDD Pacing. Pacing and Clinical Electrophysiology, 21(11), 2246–2248. https://doi.org/10.1111/j.1540-8159.1998.tb01161.x
- ISRAEL, C. W., GRÖNEFELD, G., ISCOLO, N., STÖPPLER, C., & HOHNLOSER, S. H. (2001). Discrimination Between Ventricular and Supraventricular Tachycardia by Dual Chamber Cardioverter Defibrillators: Importance of the Atrial Sensing Function. Pacing and Clinical Electrophysiology, 24(2), 183–190. https://doi.org/10.1046/j.1460-9592.2001.00183.x
Citations
- https://doi.org/10.1111/jce.14552
- https://doi.org/10.1111/pace.13256
- https://doi.org/10.1111/pace.13418
- Shurrab, M., Janmohamed, A., Sarrazin, J.-F., Ayala-Paredes, F., Sturmer, M., Williams, R., Toal, S., Lane, C., Thorpe, K. E., Healey, J. S., & Crystal, E. (2017). The Dx-AF study: a prospective, multicenter, randomized controlled trial comparing VDD-ICD to VVI-ICD in detecting sub-clinical atrial fibrillation in defibrillator patients. Journal of Interventional Cardiac Electrophysiology, 50(1), 57–63. https://doi.org/10.1007/s10840-017-0276-0
- https://doi.org/10.1111/pace.12421
- https://doi.org/10.1111/pace.12171
- https://doi.org/10.1111/j.1540-8159.2012.03452.x
- https://doi.org/10.1093/europace/eur408
- https://doi.org/10.1161/circep.110.958397
- https://doi.org/10.1161/circep.110.961219
- Rozner, M. A., & Trankina, M. (2008). Cardiac Pacing and Defibrillation. In Essentials of Cardiac Anesthesia (pp. 445–462). Elsevier. https://doi.org/10.1016/b978-141603786-6.10019-1
- https://doi.org/10.1111/j.1540-8159.2007.00681.x
- https://doi.org/10.7326/0003-4819-147-4-200708210-00007