Dead space washout by intentional leakage flow during conventional ventilation of premature infants—an experimental study
Authors/Creators
- 1. Department of Pediatric Cardiology University Hospital Erlangen Friedrich‐Alexander‐University Erlangen‐Nuremberg Erlangen Germany
-
2.
Medizinische Hochschule Hannover
-
3.
Paracelsus Medical University
Description
AbstractObjectiveInvasive mechanical ventilation poses a strong risk factor for the development of chronic lung disease in preterm infants. A reduction of the dead space as part of the total breathing volume would reduce the ventilation effort and thereby lower the risk of ventilator‐induced lung injuries. In this experimental study, we compared the efficacy of mechanical dead space washout via uncontrolled and controlled leakage flow in their ability to eliminate CO2 during conventional ventilation in preterm infants.MethodsThree frequently used neonatal ventilators, operating under standard conventional ventilating parameters, were individually connected to a test lung. To maintain a constant physiological end‐expiratory pCO2 level during ventilation, the test lung was continuously flooded with CO2. A side port in the area of the connector between the endotracheal tube and the flow sensor allowed breathing gas to escape passively or in a second experimental setup, regulated by a pump. Measurements of end‐expiratory pCO2 were taken in both experiments and compared to end‐expiratory pCO2 levels of ventilation without active dead space leakage.ResultsFollowing dead space washout, a significant reduction of end‐expiratory pCO2 was attained. Under conditions of uncontrolled leakage, the mean decrease was 14.1% while controlled leakage saw a mean reduction of 16.1%.ConclusionWashout of dead space by way of leakage flow is an effective method to reduce end‐expiratory pCO2. Both controlled and uncontrolled leakage provide comparable results, but precise regulation of leakage allows for a more stable ventilation by preventing uncontrolled loss of tidal volume during inspiration.
Additional details
Identifiers
- OpenAlex
- W4221047348
- PMID
- 35355449
- URL
- https://onlinelibrary.wiley.com/doi/10.1002/ppul.25906
Dates
- Issued
-
June 20, 2022
References
- Wald, M., Kalous, P., Lawrenz, K., Jeitler, V., Weninger, M., & Kirchner, L. (2005). Dead-space washout by split-flow ventilation. A new method to reduce ventilation needs in premature infants. Intensive Care Medicine, 31(5), 674–679. https://doi.org/10.1007/s00134-005-2611-7
- Nassabeh‐Montazami, S., Abubakar, K. M., & Keszler, M. (2008). The impact of instrumental dead‐space in volume‐targeted ventilation of the extremely low birth weight (ELBW) infant. Pediatric Pulmonology, 44(2), 128–133. https://doi.org/10.1002/ppul.20954
- Hurley, E. H., & Keszler, M. (2016). Effect of inspiratory flow rate on the efficiency of carbon dioxide removal at tidal volumes below instrumental dead space. Archives of Disease in Childhood - Fetal and Neonatal Edition, 102(2), F126–F130. https://doi.org/10.1136/archdischild-2015-309636
- Keszler, M., Montaner, M. B., & Abubakar, K. (2011). Effective ventilation at conventional rates with tidal volume below instrumental dead space: a bench study. Archives of Disease in Childhood - Fetal and Neonatal Edition, 97(3), F188–F192. https://doi.org/10.1136/archdischild-2011-300647
- Principi, N., Di Pietro, G. M., & Esposito, S. (2018). Bronchopulmonary dysplasia: clinical aspects and preventive and therapeutic strategies. Journal of Translational Medicine, 16(1). https://doi.org/10.1186/s12967-018-1417-7
- Dreyfuss, D., Soler, P., Basset, G., & Saumon, G. (1988). High Inflation Pressure Pulmonary Edema: Respective Effects of High Airway Pressure, High Tidal Volume, and Positive End-expiratory Pressure. American Review of Respiratory Disease, 137(5), 1159–1164. https://doi.org/10.1164/ajrccm/137.5.1159
- Estay, A., Claure, N., D'Ugard, C., Organero, R., & Bancalari, E. (2009). Effects of instrumental dead space reduction during weaning from synchronized ventilation in preterm infants. Journal of Perinatology, 30(7), 479–483. https://doi.org/10.1038/jp.2009.187
- Claure, N., D'Ugard, C., & Bancalari, E. (2003). Elimination of ventilator dead space during synchronized ventilation in premature infants. The Journal of Pediatrics, 143(3), 315–320. https://doi.org/10.1067/s0022-3476(03)00299-3
- Dassieu, G., Brochard, L., Agudze, E., Patkaï, J., Janaud, J.-C., & Danan, C. (1998). Continuous tracheal gas insufflation enables a volume reduction strategy in hyaline membrane disease: technical aspects and clinical results. Intensive Care Medicine, 24(10), 1076–1082. https://doi.org/10.1007/s001340050719
- Wald, M., Jeitler, V., Lawrenz, K., Weninger, M., & Kirchner, L. (2005). Effect of the Y-piece of the ventilation circuit on ventilation requirements in extremely low birth weight infants. Intensive Care Medicine, 31(8), 1095–1100. https://doi.org/10.1007/s00134-005-2711-4
- Wald, M., Jeitler, V., Lawrenz, K., Weninger, M., Pollak, A., & Kirchner, L. (2006). A Flow Sensor Suitable for Use With Split‐flow Ventilation—First Preclinical Data. Artificial Organs, 30(11), 888–891. https://doi.org/10.1111/j.1525-1594.2006.00317.x
Citations
- https://doi.org/10.3233/thc-230195
- Danan, C., Tauzin, M., Jung, C., Carbonnier, B., Dassieu, G., Decobert, F., & Caeymaex, L. (2023). Instrumental dead space: A glass ceiling for extremely low birth weight preterm infants? A dead space washout bench study. Pediatric Pulmonology, 58(5), 1514–1519. https://doi.org/10.1002/ppul.26353