Improving Real-Time Feedback During Cochlear Implantation: The Auditory Nerve Neurophonic/Cochlear Microphonic Ratio

Description

Objectives: Real-time monitoring of cochlear function to predict the loss of residual hearing after cochlear implantation is now possible. Current approaches monitor the cochlear microphonic (CM) during implantation from the electrode at the tip of the implant. A drop in CM response of >30% is associated with poorer hearing outcomes. However, there is prior evidence that CM amplitude can fluctuate in a manner unrelated to hearing trauma, leading to false positives. By monitoring another cochlear response, the auditory nerve neurophonic (ANN), a differentiation between CM drops that result in reduced cochlear output from false positives may be possible. The hypothesis tested in the present work was that ANN/CM ratios measured during a CM drop will increase during drops not associated with postoperative hearing loss. Design: Twenty-eight adult participants with known CM drops during implantation were taken from a larger data set. This contains adult cochlear implant candidates scheduled to receive a Cochlear Nucleus cochlear implant with either the slim-straight or slim-modiolar electrode array with preoperative audiometric low-frequency average thresholds of ≤80 dB HL at 500, 750, and 1000 Hz in the ear to be implanted. Patients were recruited from eight international implant sites. Pure-tone audiometry was measured postoperatively and 4 to 6 weeks after implantation. Electrocochleography was measured during and immediately after the implantation of the array in response to a 500-Hz, 6-msec pure-tone pip at 110 dB HL. Results: The ANN/CM ratio rose during CM drops in 19 of these patients and decreased in 9. At the follow-up timepoint, patients with a decreasing ANN/CM ratio had a median hearing loss of 29.0 dB, significantly worse than the group with increasing ratio at 13.3 dB (p = 0.004). Considering only the change in ANN amplitude during a CM drop led to smaller groups (ANN drop during CM drop N = 17, ANN increasing during CM drop N = 6) due to 5 patients having undetectable ANN during the CM drop. Using the ANN alone also led to as poorer prediction of hearing preservation, with median hearing preservation in the ANN increasing group of 12.9 dB, significantly better than the ANN decreasing group of 25 dB (p = 0.02). The group with a decreasing ANN/CM ratio had maximum CM amplitude immediately after insertion lower than the maximum amplitude reached during insertion (mean maximum postinsertion amplitude of 98% of during-insertion amplitude). In comparison, the ANN/CM ratio increasing group tended to have a larger CM amplitude immediately after insertion (mean maximum CM amplitude postinsertion of 164% of the maximum during-insertion amplitude). Conclusions: These data show that the ANN/CM ratio is a measure that can differentiate between patients with CM drops that lead to a loss of residual hearing and those that do not. The ANN/CM ratio is easily measured and responds rapidly during a CM drop, showing clinical promise for improving current and developing approaches to intraoperative monitoring.

Additional details

Dates

Issued
January 10, 2025

References

  1. Bester, C., Collins, A., Razmovski, T., Weder, S., Briggs, R. J., Wei, B., Zakaria, A. F., Gerard, J.-M., Mitchell-Innes, A., Tykocinski, M., Kennedy, R., Iseli, C., Dahm, M., Ellul, S., & O'Leary, S. (2022). Electrocochleography triggered intervention successfully preserves residual hearing during cochlear implantation: Results of a randomised clinical trial. Hearing Research, 426, Article 108353. https://doi.org/10.1016/j.heares.2021.108353
  2. Campbell, L., Bester, C., Iseli, C., Sly, D., Dragovic, A., Gummer, A. W., & O'Leary, S. (2017). Electrophysiological Evidence of the Basilar-Membrane Travelling Wave and Frequency Place Coding of Sound in Cochlear Implant Recipients. Audiology and Neurotology, 22(3), 180–189. https://doi.org/10.1159/000478692
  3. Campbell, L., Kaicer, A., Sly, D., Iseli, C., Wei, B., Briggs, R., & O'Leary, S. (2016). Intraoperative Real-time Cochlear Response Telemetry Predicts Hearing Preservation in Cochlear Implantation. Otology & Neurotology, 37(4), 332–338. https://doi.org/10.1097/mao.0000000000000972
  4. Dalbert, A., Pfiffner, F., Röösli, C., Thoele, K., Sim, J. H., Gerig, R., & Huber, A. M. (2015). Extra- and Intracochlear Electrocochleography in Cochlear Implant Recipients. Audiology and Neurotology, 20(5), 339–348. https://doi.org/10.1159/000438742
  5. Fontenot, T. E., Giardina, C. K., & Fitzpatrick, D. C. (2017). A Model-Based Approach for Separating the Cochlear Microphonic from the Auditory Nerve Neurophonic in the Ongoing Response Using Electrocochleography. Frontiers in Neuroscience, 11. https://doi.org/10.3389/fnins.2017.00592
  6. Forgues, M., Koehn, H. A., Dunnon, A. K., Pulver, S. H., Buchman, C. A., Adunka, O. F., & Fitzpatrick, D. C. (2014). Distinguishing hair cell from neural potentials recorded at the round window. Journal of Neurophysiology, 111(3), 580–593. https://doi.org/10.1152/jn.00446.2013
  7. Gfeller, K. E., Olszewski, C., Turner, C., Gantz, B., & Oleson, J. (2006). Music Perception with Cochlear Implants and Residual Hearing. Audiology and Neurotology, 11(Suppl. 1), 12–15. https://doi.org/10.1159/000095608
  8. Giardina, C. K., Brown, K. D., Adunka, O. F., Buchman, C. A., Hutson, K. A., Pillsbury, H. C., & Fitzpatrick, D. C. (2018). Intracochlear Electrocochleography: Response Patterns During Cochlear Implantation and Hearing Preservation. Ear & Hearing, 40(4), 833–848. https://doi.org/10.1097/aud.0000000000000659
  9. Gifford, R. H., Dorman, M. F., Skarzynski, H., Lorens, A., Polak, M., Driscoll, C. L. W., Roland, P., & Buchman, C. A. (2013). Cochlear Implantation With Hearing Preservation Yields Significant Benefit for Speech Recognition in Complex Listening Environments. Ear & Hearing, 34(4), 413–425. https://doi.org/10.1097/aud.0b013e31827e8163
  10. Golub, J. S., Won, J. H., Drennan, W. R., Worman, T. D., & Rubinstein, J. T. (2012). Spectral and Temporal Measures in Hybrid Cochlear Implant Users. Otology & Neurotology, 33(2), 147–153. https://doi.org/10.1097/mao.0b013e318241b6d3
  11. Harris, M. S., Riggs, W. J., Giardina, C. K., O'Connell, B. P., Holder, J. T., Dwyer, R. T., Koka, K., Labadie, R. F., Fitzpatrick, D. C., & Adunka, O. F. (2017). Patterns Seen During Electrode Insertion Using Intracochlear Electrocochleography Obtained Directly Through a Cochlear Implant. Otology & Neurotology, 38(10), 1415–1420. https://doi.org/10.1097/mao.0000000000001559
  12. Incerti, P. V., Ching, T. Y. C., & Cowan, R. (2013). A Systematic Review of Electric-Acoustic Stimulation. Trends in Amplification, 17(1), 3–26. https://doi.org/10.1177/1084713813480857
  13. Kiefer, J., Böhnke, F., Adunka, O., & Arnold, W. (2006). Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode. Hearing Research, 221(1-2), 36–43. https://doi.org/10.1016/j.heares.2006.07.013
  14. Lenarz, T., Buechner, A., Gantz, B., Hansen, M., Tejani, V. D., Labadie, R., O'Connell, B., Buchman, C. A., Valenzuela, C. V., Adunka, O. F., Harris, M. S., Riggs, W. J., Fitzpatrick, D., & Koka, K. (2021). Relationship Between Intraoperative Electrocochleography and Hearing Preservation. Otology & Neurotology, 43(1), e72–e78. https://doi.org/10.1097/mao.0000000000003403
  15. Lenarz, T., James, C., Cuda, D., Fitzgerald O'Connor, A., Frachet, B., Frijns, J. H. M., Klenzner, T., Laszig, R., Manrique, M., Marx, M., Merkus, P., Mylanus, E. A. M., Offeciers, E., Pesch, J., Ramos-Macias, A., Robier, A., Sterkers, O., & Uziel, A. (2013). European multi-centre study of the Nucleus Hybrid L24 cochlear implant. International Journal of Audiology, 52(12), 838–848. https://doi.org/10.3109/14992027.2013.802032
  16. Loiselle, L. H., Dorman, M. F., Yost, W. A., Cook, S. J., & Gifford, R. H. (2016). Using ILD or ITD Cues for Sound Source Localization and Speech Understanding in a Complex Listening Environment by Listeners With Bilateral and With Hearing-Preservation Cochlear Implants. Journal of Speech, Language, and Hearing Research, 59(4), 810–818. https://doi.org/10.1044/2015_jslhr-h-14-0355
  17. O'Leary, S., Briggs, R., Gerard, J.-M., Iseli, C., Wei, B. P. C., Tari, S., Rousset, A., & Bester, C. (2020). Intraoperative Observational Real-time Electrocochleography as a Predictor of Hearing Loss After Cochlear Implantation: 3 and 12 Month Outcomes. Otology & Neurotology, 41(9), 1222–1229. https://doi.org/10.1097/mao.0000000000002773
  18. O'Leary, S., Mylanus, E., Venail, F., Lenarz, T., Birman, C., Di Lella, F., Roland, J. T., Gantz, B., Beynon, A., Sicard, M., Buechner, A., Lai, W. K., Boccio, C., Choudhury, B., Tejani, V. D., Plant, K., English, R., Arts, R., & Bester, C. (2022). Monitoring Cochlear Health With Intracochlear Electrocochleography During Cochlear Implantation: Findings From an International Clinical Investigation. Ear & Hearing, 44(2), 358–370. https://doi.org/10.1097/aud.0000000000001288
  19. Palmer, A. R., & Russell, I. J. (1986). Phase-locking in the cochlear nerve of the guinea-pig and its relation to the receptor potential of inner hair-cells. Hearing Research, 24(1), 1–15. https://doi.org/10.1016/0378-5955(86)90002-x
  20. Pappa, A. K., Hutson, K. A., Scott, W. C., Wilson, J. D., Fox, K. E., Masood, M. M., Giardina, C. K., Pulver, S. H., Grana, G. D., Askew, C., & Fitzpatrick, D. C. (2019). Hair cell and neural contributions to the cochlear summating potential. Journal of Neurophysiology, 121(6), 2163–2180. https://doi.org/10.1152/jn.00006.2019
  21. Patuzzi, R. B., Yates, G. K., & Johnstone, B. M. (1989). Changes in cochlear microphonic and neural sensitivity produced by acoustic trauma. Hearing Research, 39(1-2), 189–202. https://doi.org/10.1016/0378-5955(89)90090-7
  22. Roland, J. T., Gantz, B. J., Waltzman, S. B., Parkinson, A. J., & Group, T. M. C. T. (2015). United States multicenter clinical trial of the cochlear nucleus hybrid implant system. The Laryngoscope, 126(1), 175–181. https://doi.org/10.1002/lary.25451
  23. Ruben, R. J., Bordley, J. E., & Lieberman, A. T. (1961). Cochlear potentials in man. The Laryngoscope, 71(10), 1141–1164. https://doi.org/10.1288/00005537-196110000-00001
  24. Saoji, A. A., Graham, M., Stein, A., & Koka, K. (2021). Analysis of electrode impedance and its subcomponents for lateral wall, mid-scala, and perimodiolar electrodes in cochlear implants. Cochlear Implants International, 23(2), 87–94. https://doi.org/10.1080/14670100.2021.2000734
  25. Saoji, A. A., Patel, N. S., Carlson, M. L., Neff, B. A., Koka, K., Tarigoppula, V. S. A., & Driscoll, C. L. W. (2019). Multi-frequency Electrocochleography Measurements can be Used to Monitor and Optimize Electrode Placement During Cochlear Implant Surgery. Otology & Neurotology, 40(10), 1287–1291. https://doi.org/10.1097/mao.0000000000002406

1–25 of 34