Experimental Analysis of Model-Based Roentgen Stereophotogrammetric Analysis (MBRSA) on Four Typical Prosthesis Components
Authors/Creators
-
1.
Medizinische Hochschule Hannover
- 2. Department of Orthopaedics and Division of Image Processing, and Department of Radiology, Leiden University Medical Center, P.O. Box 9600, 2300 RC Leiden, Netherlands
Description
Classical marker-based roentgen stereophotogrammetric analysis (RSA) is an accurate method of measuring in vivo implant migration. A disadvantage of the method is the necessity of placing tantalum markers on the implant, which constitutes additional manufacturing and certification effort. Model-based RSA (MBRSA) is a method by which pose-estimation of geometric surface-models of the implant is used to detect implant migration. The placement of prosthesis markers is thus no longer necessary. The accuracy of the pose-estimation algorithms used depends on the geometry of the prosthesis as well as the accuracy of the surface models used. The goal of this study was thus to evaluate the experimental accuracy and precision of the MBRSA method for four different, but typical prosthesis geometries, that are commonly implanted. Is there a relationship existing between the accuracy of MBRSA and prosthesis geometries? Four different prosthesis geometries were investigated: one femoral and one tibial total knee arthroplasty (TKA) component and two different femoral stem total hip arthroplasty (THA) components. An experimental phantom model was used to simulate two different implant migration protocols, whereby the implant was moved relative to the surrounding bone (relative prosthesis-bone motion (RM)), or, similar to the double-repeated measures performed to assess accuracy clinically, both the prosthesis and the surrounding bone model (zero relative prosthesis-bone motion (ZRM)) were moved. Motions were performed about three translational and three rotational axes, respectively. The maximum 95% confidence interval (CI) for MBRSA of all four prosthesis investigated was better than −0.034 to 0.107 mm for in-plane and −0.217 to 0.069 mm for out-of-plane translation, and from −0.038 deg to 0.162 deg for in-plane and from −1.316 deg to 0.071 deg for out-of-plane rotation, with no clear differences between the ZRM and RM protocols observed. Accuracy in translation was similar between TKA and THA components, whereas rotational accuracy about the long axis of the hip stem THA components was worse than the TKA components. The data suggest that accuracy and precision of MBRSA seem to be equivalent to the classical marker-based RSA method, at least for the nonsymmetric implant geometries investigated in this study. The model-based method thus allows the accurate measurement of implant migration without requiring prosthesis markers, and thus presents new opportunities for measuring implant migration where financial or geometric considerations of marker placement have thus far been prohibitive factors.
Additional details
Identifiers
Dates
- Issued
-
February 2, 2009
References
- Kärrholm, J. (1989). Roentgen stereophotogrammetry: Review of orthopedic applications. Acta Orthopaedica Scandinavica, 60(4), 491–503. https://doi.org/10.3109/17453678909149328
- Kärrholm, J., Hourigan, P., Timperley, J., & Razaznejad, R. (2006). Mixing bone graft with OP-1 does not improve cup or stem fixation in revision surgery of the hip: 5-year follow-up of 10 acetabular and 11 femoral study cases and 40 control cases. Acta Orthopaedica, 77(1), 39–48. https://doi.org/10.1080/17453670610045687
- Roentgen Stereophotogrammetry. A Method for the Study of the Kinematics of the Skeletal System
- Valstar, E. R., Nelissen, R. G. H. H., Reiber, J. H. C., & Rozing, P. M. (2002). The use of Roentgen stereophotogrammetry to study micromotion of orthopaedic implants. ISPRS Journal of Photogrammetry and Remote Sensing, 56(5-6), 376–389. https://doi.org/10.1016/s0924-2716(02)00064-3
- Valstar, E. R., Gill, R., Ryd, L., Flivik, G., Börlin, N., & Kärrholm, J. (2005). Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthopaedica, 76(4), 563–572. https://doi.org/10.1080/17453670510041574
- Does Early Micromotion of Femoral Stem Prostheses Matter? 4–7-Year Stereoradiographic Follow-Up of 84 Cemented Prostheses
- Roentgen Stereophotogrammetric Analysis as a Predictor of Mechanical Loosening of Knee Prostheses
- Early Migration and Late Aseptic Failure of Proximal Femoral Prostheses
- Krismer, M., Biedermann, R., Stöckl, B., Fischer, M., Bauer, R., & Haid, C. (1999). The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. The Journal of Bone and Joint Surgery. British Volume, 81-B(2), 273–280. https://doi.org/10.1302/0301-620x.81b2.0810273
- Sundfeldt, M., V Carlsson, L., B Johansson, C., Thomsen, P., & Gretzer, C. (2006). Aseptic loosening, not only a question of wear: A review of different theories. Acta Orthopaedica, 77(2), 177–197. https://doi.org/10.1080/17453670610045902
- Kaptein, B. L., Valstar, E. R., Stoel, B. C., Rozing, P. M., & Reiber, J. H. C. (2003). A new model-based RSA method validated using CAD models and models from reversed engineering. Journal of Biomechanics, 36(6), 873–882. https://doi.org/10.1016/s0021-9290(03)00002-2
- Ilchmann, T., Franzén, H., Mjöberg, B., & Wingstrand, H. (1992). Measurement accuracy in acetabular cup migration. The Journal of Arthroplasty, 7(2), 121–127. https://doi.org/10.1016/0883-5403(92)90004-a
- Biedermann, R., Krismer, M., Stöckl, B., Mayrhofer, P., Ornstein, E., & Franzén, H. (1999). Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. The Journal of Bone and Joint Surgery. British Volume, 81-B(2), 266–272. https://doi.org/10.1302/0301-620x.81b2.0810266
- Biedermann, R., Stöckl, B., Krismer, M., Mayrhofer, P., Ornstein, E., & Franzén, H. (2001). Evaluation of accuracy and precision of bone markers for the measurement of migration of hip prostheses. The Journal of Bone and Joint Surgery. British Volume, 83-B(5), 767–771. https://doi.org/10.1302/0301-620x.83b5.0830767
- Valstar, E. R., de Jong, F. W., Vrooman, H. A., Rozing, P. M., & Reiber, J. H. C. (2001). Model-based Roentgen stereophotogrammetry of orthopaedic implants. Journal of Biomechanics, 34(6), 715–722. https://doi.org/10.1016/s0021-9290(01)00028-8
- Valstar, E. R., Spoor, C. W., Nelissen, R. G. H. H., & Rozing, P. M. (1997). Roentgen stereophotogrammetric analysis of metal‐backed hemispherical cups without attached markers. Journal of Orthopaedic Research, 15(6), 869–873. https://doi.org/10.1002/jor.1100150612
- Börlin, N., Röhrl, S. M., & Bragdon, C. R. (2006). RSA wear measurements with or without markers in total hip arthroplasty. Journal of Biomechanics, 39(9), 1641–1650. https://doi.org/10.1016/j.jbiomech.2005.05.004
- Kaptein, B. L., Valstar, E. R., Spoor, C. W., Stoel, B. C., & Rozing, P. M. (2006). Model-based RSA of a Femoral Hip Stem Using Surface and Geometrical Shape Models. Clinical Orthopaedics and Related Research, 448, 92–97. https://doi.org/10.1097/01.blo.0000224010.04551.14
- Kaptein, B. L., R Valstar, E., Stoel, B. C., Rozing, P. M., & Reiber, J. H. C. (2004). Evaluation of three pose estimation algorithms for model-based Roentgen stereophotogrammetric analysis. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 218(4), 231–238. https://doi.org/10.1243/0954411041561036
- Pose Estimation of Artificial Knee Implants in Fluoroscopy Images Using a Template Matching Technique
- Mahfouz, M. R., Hoff, W. A., Komistek, R. D., & Dennis, D. A. (2005). Effect of segmentation errors on 3D-to-2D registration of implant models in X-ray images. Journal of Biomechanics, 38(2), 229–239. https://doi.org/10.1016/j.jbiomech.2004.02.025
- Dennis, D. A., Mahfouz, M. R., Komistek, R. D., & Hoff, W. (2005). In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics. Journal of Biomechanics, 38(2), 241–253. https://doi.org/10.1016/j.jbiomech.2004.02.042
- Fregly, B. J., Rahman, H. A., & Banks, S. A. (2005). Theoretical Accuracy of Model-Based Shape Matching for Measuring Natural Knee Kinematics with Single-Plane Fluoroscopy. Journal of Biomechanical Engineering, 127(4), 692–699. https://doi.org/10.1115/1.1933949
- Valstar, E. R., Vrooman, H. A., Toksvig-Larsen, S., Ryd, L., & Nelissen, R. G. H. H. (2000). Digital automated RSA compared to manually operated RSA. Journal of Biomechanics, 33(12), 1593–1599. https://doi.org/10.1016/s0021-9290(00)00138-x
- Fleming, B. C., Peura, G. D., Abate, J. A., & Beynnon, B. D. (2001). Accuracy and repeatability of Roentgen stereophotogrammetric analysis (RSA) for measuring knee laxity in longitudinal studies. Journal of Biomechanics, 34(10), 1355–1359. https://doi.org/10.1016/s0021-9290(01)00087-2
1–25 of 37
Citations
- Kistler, M., Schonhoff, M., Reulbach, M., Schleifenbaum, S., Hölscher-Doht, S., Hurschler, C., & Jahnke, A. (2024). Implantatverankerung: In-vivo- und In-vitro-Analysen: Cluster Implantatverankerung und -sicherheit. Die Orthopädie, 53(7), 519–526. https://doi.org/10.1007/s00132-024-04519-1
- Seehaus, F., Sonntag, R., Schwarze, M., Jakubowitz, E., Sesselmann, S., Kretzer, J. P., & Hurschler, C. (2020). Früherkennung des Risikos der späteren Implantatlockerung mittels der Röntgen Stereophotogrammetrischen Analyse (RSA). Der Orthopäde, 49(12), 1042–1048. https://doi.org/10.1007/s00132-020-04027-y
- Keller, M. C., Hurschler, C., & Schwarze, M. (2020). Experimental evaluation of precision and accuracy of RSA in the lumbar spine. European Spine Journal, 30(7), 2060–2068. https://doi.org/10.1007/s00586-020-06672-5
- Zinno, R., Di Paolo, S., Ambrosino, G., Alesi, D., Zaffagnini, S., Barone, G., & Bragonzoni, L. (2020). Migration of the femoral component and clinical outcomes after total knee replacement: a narrative review. MUSCULOSKELETAL SURGERY, 105(3), 235–246. https://doi.org/10.1007/s12306-020-00690-8
- Xu, J., Cao, H., Sesselmann, S., Taylor, D., Forst, R., & Seehaus, F. (2020). Model-Based Roentgen Stereophotogrammetric Analysis Using Elementary Geometrical Shape Models: Reliability of Migration Measurements for an Anatomically Shaped Femoral Stem Component. Applied Sciences, 10(23), 8507. https://doi.org/10.3390/app10238507
- Xu, J., Sonntag, R., Kretzer, J. P., Taylor, D., Forst, R., & Seehaus, F. (2020). Model-Based Roentgen Stereophotogrammetric Analysis to Monitor the Head–Taper Junction in Total Hip Arthroplasty in Vivo—And They Do Move. Materials, 13(7), 1543. https://doi.org/10.3390/ma13071543
- Van de Kleut, M. L., Yuan, X., Athwal, G. S., & Teeter, M. G. (2018). Validation of radiostereometric analysis in six degrees of freedom for use with reverse total shoulder arthroplasty. Journal of Biomechanics, 68, 126–131. https://doi.org/10.1016/j.jbiomech.2017.12.027
- https://doi.org/10.1177/0954411918784086
- Jacobsen, A., Seehaus, F., Hong, Y., Cao, H., Schuh, A., Forst, R., & Sesselmann, S. (2018). Model-based roentgen stereophotogrammetric analysis using elementary geometrical shape models: 10 years results of an uncemented acetabular cup component. BMC Musculoskeletal Disorders, 19(1). https://doi.org/10.1186/s12891-018-2259-4
- Shareghi, B., Johanson, P., & Kärrholm, J. (2017). Clinical evaluation of model‐based radiostereometric analysis to measure femoral head penetration and cup migration in four different cup designs. Journal of Orthopaedic Research, 35(4), 760–767. https://doi.org/10.1002/jor.23177
- Seehaus, F., Schwarze, M., Flörkemeier, T., von Lewinski, G., Kaptein, B. L., Jakubowitz, E., & Hurschler, C. (2016). Use of single-representative reverse-engineered surface-models for RSA does not affect measurement accuracy and precision: REVERSE-ENGINEERED MODELS FOR RSA. Journal of Orthopaedic Research, 34(5), 903–910. https://doi.org/10.1002/jor.23098
- Budde, S., Seehaus, F., Schwarze, M., Hurschler, C., Floerkemeier, T., Windhagen, H., Noll, Y., Ettinger, M., & Thorey, F. (2015). Analysis of migration of the Nanos® short-stem hip implant within two years after surgery. International Orthopaedics, 40(8), 1607–1614. https://doi.org/10.1007/s00264-015-2999-9
- Gascoyne, T. C., Morrison, J. B., & Turgeon, T. R. (2014). The effect of patient positioning on the precision of model-based radiostereometric analysis. Journal of Biomechanics, 47(8), 1914–1917. https://doi.org/10.1016/j.jbiomech.2014.03.034
- Torry, M. R., Shelburne, K. B., Myers, C., Giphart, J. E., Pennington, W. W., Krong, J. P., Peterson, D. S., Steadman, J. R., & Woo, S. L. ‐Y . (2012). High knee valgus in female subjects does not yield higher knee translations during drop landings: A biplane fluoroscopic study. Journal of Orthopaedic Research, 31(2), 257–267. https://doi.org/10.1002/jor.22217
- Tersi, L., Barré, A., Fantozzi, S., & Stagni, R. (2012). In vitro quantification of the performance of model-based mono-planar and bi-planar fluoroscopy for 3D joint kinematics estimation. Medical & Biological Engineering & Computing, 51(3), 257–265. https://doi.org/10.1007/s11517-012-0987-4
- Seehaus, F., Olender, G. D., Kaptein, B. L., Ostermeier, S., & Hurschler, C. (2012). Markerless Roentgen Stereophotogrammetric Analysis for in vivo implant migration measurement using three dimensional surface models to represent bone. Journal of Biomechanics, 45(8), 1540–1545. https://doi.org/10.1016/j.jbiomech.2012.03.004
- Syu, C.-B., Lin, S.-C., Huang, C.-Y., Lai, J.-Y., Shih, K.-S., & Chen, K.-J. (2012). Improved execution efficiency of model-based roentgen stereophotogrammetric analysis: simplification and segmentation of model meshes. Computer Methods in Biomechanics and Biomedical Engineering, 15(12), 1347–1357. https://doi.org/10.1080/10255842.2011.597352
- Dunbar, M. J., Fong, J. W., Wilson, D. A., Hennigar, A. W., Francis, P. A., & Glazebrook, M. A. (2012). Longitudinal migration and inducible displacement of the Mobility Total Ankle System. Acta Orthopaedica, 83(4), 394–400. https://doi.org/10.3109/17453674.2012.712890
- https://doi.org/10.1115/1.4005471
- https://doi.org/10.3109/17453674.2011.566133
- https://doi.org/10.3113/fai.2011.1155
- Torry, M. R., Myers, C., Pennington, W. W., Shelburne, K. B., Krong, J. P., Giphart, J. E., Steadman, J. R., & Woo, S. L. (2010). Relationship of anterior knee laxity to knee translations during drop landings: a bi‐plane fluoroscopy study. Knee Surgery, Sports Traumatology, Arthroscopy, 19(4), 653–662. https://doi.org/10.1007/s00167-010-1327-6
- Hansen, T. B., Larsen, K., Bjergelund, L., & Stilling, M. (2010). Trapeziometacarpal joint implants can be evaluated by roentgen stereophotogrammetric analysis. Journal of Hand Surgery (European Volume), 35(6), 480–485. https://doi.org/10.1177/1753193409355627