Image Quality and Radiation Exposure in Abdominal Angiography
Description
Objectives This phantom and animal pilot study aimed to compare image quality and radiation exposure between detector-dose–driven exposure control (DEC) and contrast-to-noise ratio (CNR)–driven exposure control (CEC) as functions of source-to-image receptor distance (SID) and collimation. Materials and Methods First, an iron foil simulated a guide wire in a stack of polymethyl methacrylate and aluminum plates representing patient thicknesses of 15, 25, and 35 cm. Fluoroscopic images were acquired using 5 SIDs ranging from 100 to 130 cm and 2 collimations (full field of view, collimated field of view: 6 × 6 cm). The iron foil CNRs were calculated, and radiation doses in terms of air kerma rate were obtained and assessed using a multivariate regression. Second, 5 angiographic scenarios were created in 2 anesthetized pigs. Fluoroscopic images were acquired at 2 SIDs (110 and 130 cm) and both collimations. Two blinded experienced readers compared image quality to the reference image using full field of view at an SID of 110 cm. Air kerma rate was obtained and compared using t tests. Results Using DEC, both CNR and air kerma rate increased significantly at longer SID and collimation below the air kerma rate limit. When using CEC, CNR was significantly less dependent of SID, collimation, and patient thickness. Air kerma rate decreased at longer SID and tighter collimation. After reaching the air kerma rate limit, CEC behaved similarly to DEC. In the animal study using DEC, image quality and air kerma rate increased with longer SID and collimation (P < 0.005). Using CEC, image quality was not significantly different than using longer SID or tighter collimation. Air kerma rate was not significantly different at longer SID but lower using collimation (P = 0.012). Conclusions CEC maintains the image quality with varying SID and collimation stricter than DEC, does not increase the air kerma rate at longer SID and reduces it with tighter collimation. After reaching the air kerma rate limit, CEC and DEC perform similarly.
Additional details
Identifiers
- OpenAlex
- W4393191945
- PMID
- 38529924
- URL
- https://journals.lww.com/10.1097/RLI.0000000000001079
Dates
- Issued
-
March 27, 2024
References
- Neitzel, U. (1992). Grids or air gaps for scatter reduction in digital radiography: A model calculation. Medical Physics, 19(2), 475–481. https://doi.org/10.1118/1.596836
- Sorenson, J. A., & Floch, J. (1985). Scatter rejection by air gaps: An empirical model. Medical Physics, 12(3), 308–316. https://doi.org/10.1118/1.595690
- Impact of x-ray tube voltage, field size and object thickness on scattered radiation distribution in diagnostic radiology: a Monte Carlo investigation
- Werncke, T., von Falck, C., Luepke, M., Stamm, G., Wacker, F. K., & Meyer, B. C. (2015). Collimation and Image Quality of C-Arm Computed Tomography. Investigative Radiology, 50(8), 514–521. https://doi.org/10.1097/rli.0000000000000158
- Lin, P. P. (2007). The operation logic of automatic dose control of fluoroscopy system in conjunction with spectral shaping filters. Medical Physics, 34(8), 3169–3172. https://doi.org/10.1118/1.2752576
- KAROLL, M. P., MINTZER, R. A., PAUL LIN, P.-J., SIDER, L., JOHNSON, C. S., PERLMAN, S. J., & LUBBAT, T. R. (1985). Air Gap Technique for Digital Subtraction Angiography of the Extracranial Carotid Arteries. Investigative Radiology, 20(7), 742–745. https://doi.org/10.1097/00004424-198510000-00015
- Mahnken, A. H., Boullosa Seoane, E., Cannavale, A., de Haan, M. W., Dezman, R., Kloeckner, R., O’Sullivan, G., Ryan, A., & Tsoumakidou, G. (2021). CIRSE Clinical Practice Manual. CardioVascular and Interventional Radiology, 44(9), 1323–1353. https://doi.org/10.1007/s00270-021-02904-3
- Bernhardt, P., Lendl, M., & Deinzer, F. (2006). New technologies to reduce pediatric radiation doses. Pediatric Radiology, 36(S2). https://doi.org/10.1007/s00247-006-0212-4
- Dehairs, M., Bosmans, H., Desmet, W., & Marshall, N. W. (2017). Evaluation of automatic dose rate control for flat panel imaging using a spatial frequency domain figure of merit. Physics in Medicine & Biology, 62(16), 6610–6630. https://doi.org/10.1088/1361-6560/aa7a9d
- Werncke, T., Kemling, M., Tashenov, S., Hinrichs, J. B., Meine, T. C., Maschke, S. K., Kyriakou, Y., Wacker, F. K., & Meyer, B. C. (2021). Impact of a contrast-to-noise ratio driven and material specific exposure control on image quality and radiation exposure in angiography. Physics in Medicine & Biology, 66(6), 065020. https://doi.org/10.1088/1361-6560/abe83a
- Werncke, T., Kemling, M., Tashenov, S., Hinrichs, J. B., Meine, T. C., Maschke, S. K., Kyriakou, Y., Wacker, F. K., & Meyer, B. C. (2021). Comparison of a contrast‐to‐noise ratio‐driven exposure control and a regular detector dose‐driven exposure control in abdominal imaging in a clinical angiography system. Medical Physics, 48(12), 7641–7656. https://doi.org/10.1002/mp.15288
- Vogl, T. J., Alizadeh, L. S., Maeder, R., Naguib, N. N., Herrmann, E., Bickford, M. W., Burck, I., & Albrecht, M. H. (2019). Advanced Robotic Angiography Systems for Image Guidance During Conventional Transarterial Chemoembolization. Investigative Radiology, 54(3), 153–159. https://doi.org/10.1097/rli.0000000000000519
- Dehairs, M., Bosmans, H., & Marshall, N. W. (2019). Implementation of a spatio-temporal figure of merit for new automatic dose rate control regimes in dynamic x-ray imaging. Physics in Medicine & Biology, 64(4), 045001. https://doi.org/10.1088/1361-6560/aafd5c
- White, D. R., Woodard, H. Q., & Hammond, S. M. (1987). Average soft-tissue and bone models for use in radiation dosimetry. The British Journal of Radiology, 60(717), 907–913. https://doi.org/10.1259/0007-1285-60-717-907
- AAPM task group report 272: comprehensive acceptance testing and evaluation of fluoroscopy imaging systems
- Spatial frequency-dependent signal-to-noise ratio as a generalized measure of image quality
- Impact of a high aspect ratio antiscatter grid for coronary angiography imaging on imaging system dose efficiency
- Buhr, E., Günther‐Kohfahl, S., & Neitzel, U. (2003). Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image. Medical Physics, 30(9), 2323–2331. https://doi.org/10.1118/1.1598673
- Cicchetti, D. V. (1994). Guidelines, criteria, and rules of thumb for evaluating normed and standardized assessment instruments in psychology. Psychological Assessment, 6(4), 284–290. https://doi.org/10.1037/1040-3590.6.4.284
- Norman, G. (2010). Likert scales, levels of measurement and the “laws” of statistics. Advances in Health Sciences Education, 15(5), 625–632. https://doi.org/10.1007/s10459-010-9222-y
- Miller, D. L., Balter, S., Schueler, B. A., Wagner, L. K., Strauss, K. J., & Vañó, E. (2010). Clinical Radiation Management for Fluoroscopically Guided Interventional Procedures. Radiology, 257(2), 321–332. https://doi.org/10.1148/radiol.10091269
- Wagner, L. K., Archer, B. R., & Cohen, A. M. (2000). Management of Patient Skin Dose in Fluoroscopically Guided Interventional Procedures. Journal of Vascular and Interventional Radiology, 11(1), 25–33. https://doi.org/10.1016/s1051-0443(07)61274-3
- Poletti, J., & McLean, D. (2004). The effect of source to image distance on scattered radiation to the image receptor. Australasian Physics & Engineering Sciences in Medicine, 27(4), 180–188. https://doi.org/10.1007/bf03178647
- Increasing source-to-image distance to reduce radiation dose from digital radiography pelvic examinations
- Yamashita, K., Higashino, K., Hayashi, H., Hayashi, F., Fukui, Y., & Sairyo, K. (2017). Pulsation and Collimation During Fluoroscopy to Decrease Radiation. JBJS Open Access, 2(4), e0039. https://doi.org/10.2106/jbjs.oa.17.00039
1–25 of 26