Transient Noise Reduction in Cochlear Implant Users: A Multi-Band Approach
Description
A previously-tested transient noise reduction (TNR) algorithm for cochlear implant (CI) users was modified to detect and attenuate transients independently across multiple frequency-bands. Since speech and transient noise are often spectrally distinct, we hypothesized that benefits in speech intelligibility can be achieved over the earlier single- band design. Fifteen experienced CI users (49 to 72 years) were tested unilaterally using pre-processed stimuli delivered directly to a speech processor. Speech intelligibility in transient and soft stationary noise, subjective sound quality and the recognition of warning signals was investigated in three processing conditions: no TNR (TNRoff), single- band TNR (TNRsgl) and multi-band TNR (TNRmult). Notably, TNRmult improved speech reception thresholds (SRTs) in cafeteria noise and office noise by up to 3 dB over both TNRoff and TNRsgl, and yielded higher comfort and clarity ratings in cafeteria noise. Our results indicate that multi-band transient noise reduction may be advantageous compared to a single-band approach, and reveal a substantial overall potential for TNR to improve speech perception and listening comfort in CI users.
Additional details
Identifiers
- OpenAlex
- W2514921302
- PMID
- 27942372
- URL
- https://www.mdpi.com/2039-4349/6/2/154
Dates
- Issued
-
August 23, 2016
References
- Buechner, A., Brendel, M., Saalfeld, H., Litvak, L., Frohne-Buechner, C., & Lenarz, T. (2010). Results of a Pilot Study With a Signal Enhancement Algorithm for HiRes 120 Cochlear Implant Users. Otology & Neurotology, 31(9), 1386–1390. https://doi.org/10.1097/mao.0b013e3181f1cdc6
- Henderson, D., & Hamernik, R. P. (1986). Impulse noise: Critical review. The Journal of the Acoustical Society of America, 80(2), 569–584. https://doi.org/10.1121/1.394052
- An assessment of everyday noises and their annoyance
- Korhonen, P., Kuk, F., Lau, C., Keenan, D., Schumacher, J., & Nielsen, J. (2013). Effects of a Transient Noise Reduction Algorithm on Speech Understanding, Subjective Preference, and Preferred Gain. Journal of the American Academy of Audiology, 24(9), 845–858. https://doi.org/10.3766/jaaa.24.9.8
- Buechner, A., Dyballa, K.-H., Hehrmann, P., Fredelake, S., Lenarz, T., & Wanunu, M. (2014). Advanced Beamformers for Cochlear Implant Users: Acute Measurement of Speech Perception in Challenging Listening Conditions. PLoS ONE, 9(4), e95542. https://doi.org/10.1371/journal.pone.0095542
- Transient impulse control for hearing aids
- Chalupper, J., & Powers, T. A. (2007). New algorithm is designed to take the annoyance out of transient noise. The Hearing Journal, 60(7), 42. https://doi.org/10.1097/01.hj.0000281790.84712.02
- DiGiovanni, J. J., Davlin, E. A., & Nagaraj, N. K. (2011). Effects of Transient Noise Reduction Algorithms on Speech Intelligibility and Ratings of Hearing Aid Users. American Journal of Audiology, 20(2), 140–150. https://doi.org/10.1044/1059-0889(2011/10-0007)
- Keidser, G., O'Brien, A., Latzel, M., & Convery, E. (2007). Evaluation of a noise-reduction algorithm that targets non-speech transient sounds. The Hearing Journal, 60(2), 29. https://doi.org/10.1097/01.hj.0000285643.45157.35
- Liu, H., Zhang, H., Bentler, R. A., Han, D., & Zhang, L. (2012). Evaluation of a Transient Noise Reduction Strategy for Hearing Aids. Journal of the American Academy of Audiology, 23(8), 606–615. https://doi.org/10.3766/jaaa.23.8.4
- Dyballa, K.-H., Hehrmann, P., Hamacher, V., Nogueira, W., Lenarz, T., & Büchner, A. (2015). Evaluation of A Transient Noise Reduction Algorithm in Cochlear Implant Users. Audiology Research, 5(2), 116. https://doi.org/10.4081/audiores.2015.116
- Holube, I., Fredelake, S., Vlaming, M., & Kollmeier, B. (2010). Development and analysis of an International Speech Test Signal (ISTS). International Journal of Audiology, 49(12), 891–903. https://doi.org/10.3109/14992027.2010.506889
- Peterson, G. E., & Barney, H. L. (1952). Control Methods Used in a Study of the Vowels. The Journal of the Acoustical Society of America, 24(2), 175–184. https://doi.org/10.1121/1.1906875
- Transient interference suppression in speech signals based on the OM-LSA algorithm
- Brendel, M., Buechner, A., Krueger, B., Frohne-Buechner, C., & Lenarz, T. (2008). Evaluation of the Harmony Soundprocessor in Combination With the Speech Coding Strategy HiRes 120. Otology & Neurotology, 29(2), 199–202. https://doi.org/10.1097/mao.0b013e31816335c6
- Boyle, P. J., Büchner, A., Stone, M. A., Lenarz, T., & Moore, B. C. J. (2009). Comparison of dual-time-constant and fast-acting automatic gain control (AGC) systems in cochlear implants. International Journal of Audiology, 48(4), 211–221. https://doi.org/10.1080/14992020802581982
- Kollmeier, B., & Wesselkamp, M. (1997). Development and evaluation of a German sentence test for objective and subjective speech intelligibility assessment. The Journal of the Acoustical Society of America, 102(4), 2412–2421. https://doi.org/10.1121/1.419624
- The HSM sentence test as a tool for evaluating the speech understanding in noise of cochlear implant users
- Practical nonparametric statistics
- Is the hearing instrument to blame when things get really noisy