Published July 5, 2011 | Journal article | Version v1 | https://doi.org/10.1186/1475-925x-10-58

Numerical simulation of strain-adaptive bone remodelling in the ankle joint

Description

AbstractBackgroundThe use of artificial endoprostheses has become a routine procedure for knee and hip joints while ankle arthritis has traditionally been treated by means of arthrodesis. Due to its advantages, the implantation of endoprostheses is constantly increasing. While finite element analyses (FEA) of strain-adaptive bone remodelling have been carried out for the hip joint in previous studies, to our knowledge there are no investigations that have considered remodelling processes of the ankle joint. In order to evaluate and optimise new generation implants of the ankle joint, as well as to gain additional knowledge regarding the biomechanics, strain-adaptive bone remodelling has been calculated separately for the tibia and the talus after providing them with an implant.MethodsFE models of the bone-implant assembly for both the tibia and the talus have been developed. Bone characteristics such as the density distribution have been applied corresponding to CT scans. A force of 5,200 N, which corresponds to the compression force during normal walking of a person with a weight of 100 kg according to Stauffer et al., has been used in the simulation. The bone adaptation law, previously developed by our research team, has been used for the calculation of the remodelling processes.ResultsA total bone mass loss of 2% in the tibia and 13% in the talus was calculated. The greater decline of density in the talus is due to its smaller size compared to the relatively large implant dimensions causing remodelling processes in the whole bone tissue. In the tibia, bone remodelling processes are only calculated in areas adjacent to the implant. Thus, a smaller bone mass loss than in the talus can be expected. There is a high agreement between the simulation results in the distal tibia and the literature regarding.ConclusionsIn this study, strain-adaptive bone remodelling processes are simulated using the FE method. The results contribute to a better understanding of the biomechanical behaviour of the ankle joint and hence are useful for the optimisation of the implant geometry in the future.

Additional details

Dates

Issued
July 5, 2011

References

  1. Rao, S., Ellis, S. J., Deland, J. T., & Hillstrom, H. (2010). Nonmedicinal therapy in the management of ankle arthritis. Current Opinion in Rheumatology, 22(2), 223–228. https://doi.org/10.1097/bor.0b013e328335fceb
  2. Hintermann, B. (ed.). (2005). Endoprothetik des Sprunggelenks: Historischer Überblick, aktuelle Therapiekonzepte und Entwicklungen (B. Hintermann, ed.). Springer Vienna. https://doi.org/10.1007/3-211-30885-7
  3. Bauer, G., Eberhardt, O., Rosenbaum, D., & Claes, L. (1996). Total ankle replacement. Review and critical analysis of the current status. Foot and Ankle Surgery, 2(2), 119–126. https://doi.org/10.1046/j.1460-9584.1996.00024.x
  4. Doets, H. C., Middelkoop, M. van ., Houdijk, H., Nelissen, R. G. H. H., & Veeger, H. E. J. (Dirkjan) . (2007). Gait Analysis After Successful Mobile Bearing Total Ankle Replacement. Foot & Ankle International, 28(3), 313–322. https://doi.org/10.3113/fai.2007.0313
  5. Kofoed, H., & Stürup, J. (1994). Comparison of ankle arthroplasty and arthrodesis. A prospective series with long-term follow-up. The Foot, 4(1), 6–9. https://doi.org/10.1016/0958-2592(94)90003-5
  6. MCGUIRE, M. R., KYLE, R. F., GUSTILO, R. B., & PREMER, R. F. (1988). Comparative Analysis of Ankle Arthroplasty Versus Ankle Arthrodesis. Clinical Orthopaedics and Related Research, 226, 174–181. https://doi.org/10.1097/00003086-198801000-00024
  7. Gougoulias, N. E., Khanna, A., & Maffulli, N. (2008). History and evolution in total ankle arthroplasty. British Medical Bulletin, 89(1), 111–151. https://doi.org/10.1093/bmb/ldn039
  8. Vickerstaff, J. A., Miles, A. W., & Cunningham, J. L. (2007). A brief history of total ankle replacement and a review of the current status. Medical Engineering & Physics, 29(10), 1056–1064. https://doi.org/10.1016/j.medengphy.2006.11.009
  9. Leardini, A., O'Connor, J. J., Catani, F., & Giannini, S. (1999). A geometric model of the human ankle joint. Journal of Biomechanics, 32(6), 585–591. https://doi.org/10.1016/s0021-9290(99)00022-6
  10. Stauffer RN, Chao EYS, Brewster RC: Force and motion analysis of the normal, diseased, and prosthetic ankle joint. Orthopaedics and Related Research 1977, 127: 189–196.
  11. BURDETT, R. G. (1982). Forces predicted at the ankle during running. Medicine & Science in Sports & Exercise, 14(4), 308–316. https://doi.org/10.1249/00005768-198204000-00010
  12. Procter, P., & Paul, J. P. (1982). Ankle joint biomechanics. Journal of Biomechanics, 15(9), 627–634. https://doi.org/10.1016/0021-9290(82)90017-3
  13. Röhrle, H., Scholten, R., Sigolotto, C., Sollbach, W., & Kellner, H. (1984). Joint forces in the human pelvis-leg skeleton during walking. Journal of Biomechanics, 17(6), 409–424. https://doi.org/10.1016/0021-9290(84)90033-2
  14. Seireg, A., & Arvikar, R. J. (1975). The prediction of muscular load sharing and joint forces in the lower extremities during walking. Journal of Biomechanics, 8(2), 89–102. https://doi.org/10.1016/0021-9290(75)90089-5
  15. Benedetti, M., Catani, F., Leardini, A., Pignotti, E., & Giannini, S. (1998). Data management in gait analysis for clinical applications. Clinical Biomechanics, 13(3), 204–215. https://doi.org/10.1016/s0268-0033(97)00041-7
  16. Leardini, A. (2001). Geometry and mechanics of the human ankle complex and ankle prosthesis design. Clinical Biomechanics, 16(8), 706–709. https://doi.org/10.1016/s0268-0033(01)00022-5
  17. Gougoulias, N., Khanna, A., & Maffulli, N. (2010). How Successful are Current Ankle Replacements?: A Systematic Review of the Literature. Clinical Orthopaedics & Related Research, 468(1), 199–208. https://doi.org/10.1007/s11999-009-0987-3
  18. Haddad, S. L., Coetzee, J. C., Estok, R., Fahrbach, K., Banel, D., & Nalysnyk, L. (2007). Intermediate and Long-Term Outcomes of Total Ankle Arthroplasty and Ankle Arthrodesis. The Journal of Bone & Joint Surgery, 89(9), 1899–1905. https://doi.org/10.2106/jbjs.f.01149
  19. S.T.A.R. ® Ankle Implants and Instruments Guide, Small Bone Innovations 2011. [
  20. S Fevang, B.-T., A Lie, S., I Havelin, L., G Brun, J., Skredderstuen, A., & Furnes, O. (2007). 257 ankle arthroplasties performed in Norway between 1994 and 2005. Acta Orthopaedica, 78(5), 575–583. https://doi.org/10.1080/17453670710014257
  21. Henricson, A., Skoog, A., & Carlsson, Å. (2007). The Swedish Ankle Arthroplasty Register: An analysis of 531 arthroplasties between 1993 and 2005. Acta Orthopaedica, 78(5), 569–574. https://doi.org/10.1080/17453670710014248
  22. Kummer B: Biomechanik. Form und Funktion des Bewegungsapparates. Köln, Deutscher Ärzte-Verlag; 2005.
  23. Wolff J: Das Gesetz der Transformation der Knochen. Berlin, Hirschwald; 1892.
  24. Bouguecha A: Numerische und experimentelle Untersuchungen zum beanspruchungsadaptiven Knochenumbau im periprothetischen caninen Femur. Leibniz Universität Hannover, Institut für Umformtechnik und Umformmaschinen; 2007.
  25. Behrens B-A, Bouguecha A: Femorale Belastungsabschirmung nach einer Hüftarthroplastik. 57. Jahrestagung der Norddeutschen Orthopädenvereinigung, 12.-14.06. Hamburg 2008.

1–25 of 48

Citations

  1. Schuberth, J. M., & Christensen, J. C. (2025). Principles of Revision Total Ankle Arthroplasty. Clinics in Podiatric Medicine and Surgery, 42(2), 347–370. https://doi.org/10.1016/j.cpm.2024.12.001
  2. Enam, S., Hong Seng, G., & Ramlee, M. H. (2024). Patient-Specific Design of Knee and Ankle Implant: A Short Review on the Design Process, Analysis, Manufacturing, and Clinical Outcomes. Malaysian Journal of Medicine and Health Sciences, 20(2), 312–323. https://doi.org/10.47836/mjmhs.20.2.40
  3. Talbott, H., Jha, S., Gulati, A., Brockett, C., Mangwani, J., & Pegg, E. C. (2023). Clinically useful finite element models of the natural ankle – A review. Clinical Biomechanics, 106, 106006. https://doi.org/10.1016/j.clinbiomech.2023.106006
  4. Guo, Y., & Guo, W. (2022). Study and numerical analysis of Von Mises stress of a new tumor-type distal femoral prosthesis comprising a peek composite reinforced with carbon fibers: finite element analysis. Computer Methods in Biomechanics and Biomedical Engineering, 25(15), 1663–1677. https://doi.org/10.1080/10255842.2022.2032681
  5. Malherbe, C., Deleu, P.-A., Devos Bevernage, B., Birch, I., Maldague, P., Gombault, V., Putzeys, P., & Leemrijse, T. (2022). Early-Term Results of the Cadence Total Ankle Prosthesis: An European Noninventor Study. Foot & Ankle International, 44(1), 1–12. https://doi.org/10.1177/10711007221136539
  6. van Hoogstraten, S. W. G., Hermus, J., Loenen, A. C. Y., Arts, J. J., & van Rietbergen, B. (2022). Malalignment of the total ankle replacement increases peak contact stresses on the bone-implant interface: a finite element analysis. BMC Musculoskeletal Disorders, 23(1). https://doi.org/10.1186/s12891-022-05428-0
  7. Wiebking, U., Gaedke, I., O'Loughlin, F., & Gaulke, R. (2022). Bone density loss following ankle fusion persists at long-term follow-up. Archives of Trauma Research, 11(2), 59. https://doi.org/10.4103/atr.atr_3_22
  8. Venjakob, E., Yao, D., Ettinger, S., Claassen, L., Schwarze, M., Plaass, C., Stukenborg-Colsman, C., & Lerch, M. (2021). Feasibility and reliability of DEXA analysis after total ankle arthroplasty: A cadaver study. Foot and Ankle Surgery, 27(1), 110–115. https://doi.org/10.1016/j.fas.2020.03.001
  9. Dohle, J., Stukenborg-Colsman, C., & Plaaß, C. (2021). Endoprothetik am oberen Sprunggelenk. Zeitschrift für Orthopädie und Unfallchirurgie, 160(03), 341–360. https://doi.org/10.1055/a-1284-6896
  10. Bansod, Y. D., Kebbach, M., Kluess, D., Bader, R., & van Rienen, U. (2021). Computational Analysis of Bone Remodeling in the Proximal Tibia Under Electrical Stimulation Considering the Piezoelectric Properties. Frontiers in Bioengineering and Biotechnology, 9. https://doi.org/10.3389/fbioe.2021.705199
  11. Zafar, M. J., Kallemose, T., Benyahia, M., Ebskov, L. B., & Penny, J. Ø. (2020). 12-year survival analysis of 322 Hintegra total ankle arthroplasties from an independent center. Acta Orthopaedica, 91(4), 444–449. https://doi.org/10.1080/17453674.2020.1751499
  12. Mondal, S., & Ghosh, R. (2020). Influence of cancellous bone material and dead zone on stress-strain, bone stimulus and bone remodelling around the tibia for total ankle replacement. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 235(2), 185–196. https://doi.org/10.1177/0954411920967775
  13. Wang, Y., Wong, D. W.-. chi ., Tan, Q., Li, Z., & Zhang, M. (2019). Total ankle arthroplasty and ankle arthrodesis affect the biomechanics of the inner foot differently. Scientific Reports, 9(1). https://doi.org/10.1038/s41598-019-50091-6
  14. Mondal, S., & Ghosh, R. (2019). Bone remodelling around the tibia due to total ankle replacement: effects of implant material and implant–bone interfacial conditions. Computer Methods in Biomechanics and Biomedical Engineering, 22(16), 1247–1257. https://doi.org/10.1080/10255842.2019.1661385
  15. Mondal, S., & Ghosh, R. (2019). Effects of implant orientation and implant material on tibia bone strain, implant–bone micromotion, contact pressure, and wear depth due to total ankle replacement. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 233(3), 318–331. https://doi.org/10.1177/0954411918823811
  16. Wang, Y., Li, Z., Wong, D. W.-C., Cheng, C.-K., & Zhang, M. (2018). Finite element analysis of biomechanical effects of total ankle arthroplasty on the foot. Journal of Orthopaedic Translation, 12, 55–65. https://doi.org/10.1016/j.jot.2017.12.003
  17. Terrier, A., Fernandes, C. S., Guillemin, M., & Crevoisier, X. (2017). Fixed and mobile-bearing total ankle prostheses: Effect on tibial bone strain. Clinical Biomechanics, 48, 57–62. https://doi.org/10.1016/j.clinbiomech.2017.07.009
  18. Strauß, V., Scheer, A.-C., & Andermahr, J. (2017). Pathogenese von Knochenzysten nach Sprunggelenksendoprothesen. Fuß & Sprunggelenk, 15(2), 97–107. https://doi.org/10.1016/j.fuspru.2017.04.001
  19. Wang, Y., Wong, D. W.-C., & Zhang, M. (2015). Computational Models of the Foot and Ankle for Pathomechanics and Clinical Applications: A Review. Annals of Biomedical Engineering, 44(1), 213–221. https://doi.org/10.1007/s10439-015-1359-7
  20. Almohallami, A., Behrens, B.-A., Bouguecha, A., Lerch, M., Betancur, S., Stukenborg-Colsman, C., Nolte, I., & Lucas, K. (2015). Numerical analysis of the biomechanical complications accompanying the total hip replacement with NANOS-Prosthetic: bone remodelling and prosthesis migration. Current Directions in Biomedical Engineering, 1(1), 381–385. https://doi.org/10.1515/cdbme-2015-0093
  21. Latifi, H., Xie, Y. M., Huang, X., & Bilgen, M. (2014). Computational Simulations of Bone Remodeling under Natural Mechanical Loading or Muscle Malfunction Using Evolutionary Structural Optimization Method. Engineering, 06(03), 113–126. https://doi.org/10.4236/eng.2014.63015
  22. Gaden, M. T. R., & Ollivere, B. J. (2013). Periprosthetic Aseptic Osteolysis in Total Ankle Replacement. Clinics in Podiatric Medicine and Surgery, 30(2), 145–155. https://doi.org/10.1016/j.cpm.2012.10.006
  23. Fang, J., Gong, H., Kong, L., & Zhu, D. (2013). Simulation on the internal structure of three-dimensional proximal tibia under different mechanical environments. BioMedical Engineering OnLine, 12(1). https://doi.org/10.1186/1475-925x-12-130